Chemical cleaning (CLEANING option)
This option is displayed only if the
CLEANING
option has been activated.
Define the parameters required for a chemical cleaning of the system.
1.
Select
CLEANING
with the scrolling keys and press
Enter
.
Option
Description
FREQUENCY
Enter the number of measurements between each chemical cleaning.
SPRINKLER
Define the duration (in seconds) for rinsing the measurement chamber.
REAG. INJ
Define the duration (in seconds) that the cleaning solution is injected into the measurement
chamber.
RESIDENC.
Define the duration (in seconds) that the cleaning solution remains in the measurement
chamber.
Sample conditioning (SAMPLE COND option)
This option is displayed only if the
SAMPLE COND
option has been activated.
Define the parameters required for conditioning the sample prior to measurement. It is possible to
define 1 or 2 conditioning reagents.
1.
Select
SAMPLE COND
with the scrolling keys and press
Enter
.
Option
Description
REAGENT 1/2 OFF/ON
Set to
ON
if the conditioning is made using this reagent.
INJ. REAG. 1/2
Define the duration (in seconds) that the reagent is injected into the measurement
chamber.
TIME REAG. 1/2
Define the duration (in seconds) that the reagent remains in the sample before
measurement.
View calibration details (CAL PARAMETER option)
This option lets you view the details of the primary and last calibrations made by the analyzer.
1.
Select
CAL PARAMETER
with the scrolling keys and press
Enter
.
2.
Use the scroll keys to select
LAST CAL
or
PRIMARY CAL
, then scroll through the details by
pressing
Enter
.
Measurement cycle parameters (TIMING option)
Define the parameters for each measurement.
1.
Select
TIMING
with the scrolling keys and press
Enter
.
Option
Description
SAMPLE TIME
Define the time (in seconds) the sample is injected into the measuring chamber.
SPRINKLER N
Define the number of rinses required.
SPRINKLER t
Define the duration (in seconds) of each rinse.
INJECT TIME
Define the time (in seconds) the reagent is injected into the sample.
DELAY TIME
Define the time (in seconds) between each injection.