3.3 Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.

Electrical equipment marked with this symbol may not be disposed of in European
domestic or public disposal systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.

3.4 Product hazards

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.

C A U T I O N

 

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.

C A U T I O N

 

Personal injury hazard. Glass components can break. Handle with care to prevent cuts.

Section 4 Preparation for use

N O T I C E

 

Make sure to remove the protective tape from the filling hole of new probes. A probe with a blocked
filling hole will not operate correctly.

New probes are filled with electrolyte filling solution and have a soaker bottle that contains storage
solution to keep the glass bulb and reference junction hydrated. Prepare the probe for calibration and
measurement as follows.

1.

Remove the protective tape from the filling hole. Refer to 

Figure 2

.

Note: Put the soaker bottle on the soaker bottle holder when not in use.

2.

Rinse the reference junction and glass bulb with deionized water. Blot dry with a lint-free cloth.

3.

If the inner filling solution is low, add more filling solution. Refer to 

Fill the probe

 on page 12.

4.

If the level of KCl crystals in the probe is less than approximately 0.5 cm (0.2 in.), or for
measurements at high temperatures (> 60 °C (140 °F), add more KCl crystals. Refer to 

Add KCl

crystals to the probe

 on page 13.

5.

Invert the probe to make sure that the KCl crystals move freely through the liquid. If necessary,
shake the probe to loosen the crystals.

6.

Soak the probe in a warm pH 4 buffer solution for approximately 1 hour to condition the probe.

Note: If probe stabilization is slow, soak the probe for 2 hours in a warm pH 4 buffer solution. The soak
procedure helps to dissolve KCl crystals in the reference junction.

English

   

5

Содержание PHC735

Страница 1: ...DOC022 53 80434 PHC735 05 2021 Edition 2 User Manual...

Страница 2: ......

Страница 3: ...rocedure 8 Section 6 Sample measurement 9 6 1 Sample requirements 9 6 2 Sample measurement notes 9 6 3 Sample measurement procedure 9 Section 7 Verify the calibration 10 7 1 Verification procedure 10...

Страница 4: ...Table of Contents 2...

Страница 5: ...kaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 5 Probe soaker bottle holder 2 Filling hole plug 6 Electrode filling and storage solution 3 Protective tape and filli...

Страница 6: ...nufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water qu...

Страница 7: ...probe with a blocked filling hole will not operate correctly New probes are filled with electrolyte filling solution and have a soaker bottle that contains storage solution to keep the glass bulb and...

Страница 8: ...e meter starts for the first time or after battery replacement 8 Connect the probe to the meter 9 Remove the filling hole plug before use Keep the level of the electrolyte filling solution above the l...

Страница 9: ...e pH units apart For a one point calibration use a pH buffer near the expected sample pH Use the default calibration options or change the options in the probe settings menu Use the single display mod...

Страница 10: ...ffer solution Make sure that the sensor and reference junction are fully in the solution Do not put the probe on the bottom or sides of the beaker 6 Shake the probe from side to side to refresh the re...

Страница 11: ...purity or low ionic strength LIS samples absorb gases from the air that can change the sample pH For stable measurements use an LIS chamber or a flow cell 6 2 Sample measurement notes Read the notes...

Страница 12: ...res the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solution and acceptance criteria for verification in the probe specif...

Страница 13: ...ample material stays on the probe 1 Rinse the probe with deionized water Use warm 35 45 C 95 113 F deionized water to remove storage solution that dries on the probe Dry the probe body with a lint fre...

Страница 14: ...nctions Electrode cleaning solution for porous pin diaphragm reference junctions Thiourea 3 hours with stirring or until the porous plug becomes white 8 2 Fill the probe Add electrolyte filling soluti...

Страница 15: ...s less than 0 5 cm 0 2 in or when the sample temperature is more than 60 C 140 F Refer to Figure 5 and Table 2 for preparation Mix some of the filling solution with the crystals to help move the cryst...

Страница 16: ...d KCl crystals from the probe as possible Discard the solution and KCl crystals 3 Remove the remaining KCl crystals as follows a Fill the probe half full with deionized water Close the filling hole b...

Страница 17: ...N O T I C E Probes can become permanently damaged if kept in a storage solution that is not specified by the manufacturer Use only the specified storage solution saturated KCl Do not store the probe i...

Страница 18: ...ilization or inaccurate measurements The sample absorbs carbon dioxide CO2 from the air which causes the pH value to slowly decrease in low ionic strength LIS or high purity samples Use the LIS chambe...

Страница 19: ...Item no pH 4 01 buffer solution Singlet one use packets 20 mL each 20 pkg 2770020 pH 7 00 buffer solution Singlet one use packets 20 mL each 20 pkg 2770120 pH 10 01 buffer solution Singlet one use pac...

Страница 20: ...pkg SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder...

Страница 21: ......

Страница 22: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: