3.20 User mode calibration Linc
TM1
function
The
Linc
TM
function allows for calibration when the meter is in the User mode and
the user does not have a user standard solution to calibrate the meter. This ensures
more accurate measurements. When the Linc function is used, the User mode is
linked to another standard solution. For example: If User and KCl are linked, a KCl
standard solution is used to calibrate the instrument.
Note:
When a “Linc” is established for the User mode, the Linc applies to all measurement
modes using the User solution selection.
3.20.1 Calibrate meter for User mode
To calibrate the meter for the User mode:
1.
Push the
COND
key, the
MIN/SALT
key, or the
TDS
key.
2.
Calibrate the meter using a standard solution (see
).
3.
Select the User mode (see
).
4.
Set the calibration Linc.
3.20.2 Set User mode calibration Linc
The Linc function sets the calibration offset factor of a standard solution to the User
solution mode. The Linc stays intact in future calibrations until it is canceled (see
).
Follow the steps below to set the KCl, NaCl or 442 calibration factor to the User
solution mode.
1.
Push a measurement key to Linc (i.e.,
COND
,
RES
,
MIN/SALT
or
TDS
).
2.
Select the User
mode (see
).
1
Trademark of Myron L Company
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...