Operation
25
3.
Push the
CAL/MCLR
key to switch units.
4.
Push any measurement key to accept the unit preference for all temperature
readings.
Note:
Temperature compensation is always displayed in %/
º
C.
3.17 Return to factory settings
To set all calibrations to factory settings or to erase all records, follow the steps
below.
1.
Push any measurement key.
2.
Push the
MR/DOWN
key repeatedly until FAC SEL is displayed. To quickly
scroll through all stored memory records, hold down
MR/DOWN
.
3.
Push the
CAL/MCLR
key to accept the factory reset. The meter returns to the
measurement mode.
3.18 Cell check
The cell check verifies the cleanliness of the conductivity/TDS/resistivity sensor. If
the display shows
.00
when the cell cup is dry, the sensor is probably clean.
In normal use, the conductivity cell may become dirty or coated and require
cleaning. To perform a cell check:
1.
Push the
COND
key.
2.
Push the
MR/DOWN
key repeatedly until the display shows CELL ch. To
quickly scroll through all stored memory records, hold down
MR/DOWN
.
3.
Push the
CAL/MCLR
key to test. If the cell is clean, “Good” displays briefly. If
cell is dirty, “Cell cLn”
displays. To clean the sensors, (see
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...