Operation
20
3.12 Change the user-selected temperature
compensation factor
Select the User mode to change the temperature compensation factor. This feature
does not apply to pH or ORP. For user mode information, (see
).
1.
Select the User mode (see
2.
Push the
CAL/MCLR
key.
3.
Push
UP/MS
or
MR/DOWN
to adjust the temperature compensation factor
from 0-9.99%/
º
C
.
4.
Push the
CAL/MCLR
key twice to skip calibration adjustment and accept the
new temperature compensation (three times if in TDS or MIN/SALT mode).
5.
Measure samples with the new temperature compensation factor.
3.12.1 Disable temperature compensation
1.
Select the User mode (see
2.
Push the
CAL/MCLR
key. Hold the
MR/DOWN
key until the temperature
compensation shows .00%/
º
C.
3.
Push the
CAL/MCLR
key twice (three times for TDS or MIN/SALT).
4.
Temperature compensation is now disabled (=0) for measurements in the User
mode.
3.13 Change the user-selected conductivity/TDS ratio
Select the User mode to change a custom conductivity/TDS conversion ratio in the
range of 0.20 to 7.99.
To determine the conversion ratio for a custom solution of a known TDS ppm value,
measure the solution conductivity at 25 ºC with the MP Series meter and divide the
Содержание MP-4
Страница 2: ...2 ...
Страница 6: ...Table of contents 6 ...
Страница 10: ...Specifications 10 ...
Страница 30: ...Operation 30 ...
Страница 36: ...Calibration 36 ...
Страница 42: ...Troubleshooting 42 ...
Страница 48: ...Limited Warranty 48 ...
Страница 52: ...Conductivity conversion 52 ...
Страница 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...