Operation
Send data to the computer (2100Q)
Data that has been downloaded to the USB+Power module or USB
module can be transferred and stored to a computer. The data will be
formatted as an XML file.
1.
Push
DATA MANAGEMENT>Send Data Log
.
2.
Select an option.
Option
Description
Send Data
Log
All data is sent to the USB+Power/USB module.
Filter Data
Data is filtered by:
Data Type—All Logs, Reading Log, Calibration Log and
Verify Cal Log
Time Interval—All Logs, Last Reading, Current Day, Current
Week and Current Month
Sample ID—Off or Sample ID
Operator ID—Off or Operator ID
a.
Select Send Data to send all data to the USB+Power/USB
module.
b.
Select Filter Data. Make a choice of the data to transfer and push
Exit. Select Send Data to send the selected data to the USB
+Power module.
A progress bar will show the transfer status and shows "Transfer
Complete" when the data is successfully sent to the USB
+Power/USB module.
3.
Remove the USB+Power/USB module from the meter.
Note: Do not connect the USB+Power/USB module to the computer when the
module is still connected to the meter.
4.
Plug the USB cable, type A, to the USB+Power/USB module and
connect the USB+Power/USB module to the PC.
Note: The USB+Power/USB module works like a USB storage device.
Navigate through Windows Explorer to find the USB+power module.
5.
Open the folder "datalog". Find the data file.
The file name will be formatted as: Year_Month_Day_Hour_Minute.
xml.
6.
Save the data file to a location on the computer.
A progress bar will show the transfer status and shows "Transfer
Complete" when the data is successfully sent to the USB
+Power/USB module.
7.
Open a spreadsheet program to view the data.
A progress bar will show the transfer status and shows "Transfer
Complete" when the data is successfully sent to the USB
+Power/USB module.
Send data to the computer (DR 1900)
Data that has been downloaded to the USB+Power or USB module can
be transferred and stored to a computer. The data will be formatted as
an CSV file.
1.
Push
SETTINGS>Data Management>Send Data
. All data is sent to
the USB+Power/USB module.
The reading screen shows "Transfer Complete" when the data is
successfully sent to the USB+Power/USB module.
2.
Remove the USB+Power/USB module from the meter.
Note: Do not connect the USB+Power/USB module to the computer when the
module is still connected to the meter.
3.
Plug the USB cable, type A, to the USB+Power/USB module and
connect the USB+Power/USB module to the PC.
Note: The USB+Power/USB module works like a USB storage device.
Navigate through Windows Explorer to find the USB+power/USB module.
4.
The data file is located under the root directory.
The file name will be formatted as: LOG-
Year_Month_Day_Hour_Minute_Second.txt.
5.
Save the data file to a location on the computer.
6.
Open a spreadsheet program to view the data.
10
English
Содержание LANGE LZV813
Страница 7: ...English 7...
Страница 8: ...8 English...
Страница 9: ...English 9...
Страница 17: ...Deutsch 17...
Страница 18: ...18 Deutsch...
Страница 19: ...Deutsch 19...
Страница 27: ...Italiano 27...
Страница 28: ...28 Italiano...
Страница 29: ...Italiano 29...
Страница 37: ...Fran ais 37...
Страница 38: ...38 Fran ais...
Страница 39: ...Fran ais 39...
Страница 47: ...Espa ol 47...
Страница 48: ...48 Espa ol...
Страница 49: ...Espa ol 49...
Страница 57: ...Portugu s 57...
Страница 58: ...58 Portugu s...
Страница 59: ...Portugu s 59...
Страница 66: ...66 e tina...
Страница 67: ...e tina 67...
Страница 68: ...68 e tina...
Страница 76: ...76 Dansk...
Страница 77: ...Dansk 77...
Страница 78: ...78 Dansk...
Страница 86: ...86 Nederlands...
Страница 87: ...Nederlands 87...
Страница 88: ...88 Nederlands...
Страница 96: ...96 Polski...
Страница 97: ...Polski 97...
Страница 98: ...98 Polski...
Страница 105: ...Svenska 105...
Страница 106: ...106 Svenska...
Страница 107: ...Svenska 107...
Страница 115: ...Suomi 115...
Страница 116: ...116 Suomi...
Страница 117: ...Suomi 117...
Страница 121: ...121 128 121 130 123 100 240 V 50 60 Hz 9V 2A USB 4 AA NiMH USB 0 40 C 32 104 F 121...
Страница 122: ...12 2005 2002 96 2100Q 2100Q DR 1900 USB USB USB USB USB PD 24 Citizen USB LZV803 1 1 1 1 2 1 USB 2 122...
Страница 123: ...2 USB 1 USB 1 3 USB A USB 2 USB A USB 1 USB LZV803 3 3 USB 1 USB A USB 3 USB 1 2 USB A USB 4 1 NiMH 123...
Страница 124: ...NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH AA 10 30 USB USB USB 124...
Страница 125: ...125...
Страница 126: ...126...
Страница 127: ...127...
Страница 131: ...USB USB 1 LZV818 USB 1 LZV819 1 131...
Страница 136: ...136 Magyar...
Страница 137: ...Magyar 137...
Страница 138: ...138 Magyar...
Страница 146: ...146 Rom n...
Страница 147: ...Rom n 147...
Страница 148: ...148 Rom n...
Страница 152: ...152 159 152 161 154 100 240 50 60 9 2 USB 4 NiMH USB 0 40 C 32 104 F 152...
Страница 153: ...12 2005 2002 96 EC 2100Q 2100Q DR 1900 USB USB USB USB USB PD 24 Citizen USB LZV803 1 1 1 1 2 1 USB 2 153...
Страница 155: ...NiMH NiMH 1 2 1 5 NiMH 2300 90 NiMH 10 o c e 30 USB USB USB 155...
Страница 156: ...156...
Страница 157: ...157...
Страница 158: ...158...
Страница 162: ...USB 1 LZV566 110 240 1 LZV803 1 LZV824 USB 1 LZV825 1 LZV826 USB USB 1 LZV818 USB 1 LZV819 1 162...
Страница 167: ...T rk e 167...
Страница 168: ...168 T rk e...
Страница 169: ...T rk e 169...
Страница 177: ...Slovenski 177...
Страница 178: ...178 Slovenski...
Страница 179: ...Slovenski 179...
Страница 187: ...Hrvatski 187...
Страница 188: ...188 Hrvatski...
Страница 189: ...Hrvatski 189...
Страница 193: ...193 200 193 203 195 AC 100 240 V 50 60 Hz 9 V 2 A DC USB 4 NiMH USB 0 40 C 32 104 F 193...
Страница 194: ...12 2005 2002 96 2100Q 2100Qis DR 1900 USB USB AC DC USB USB USB Citizen PD 24 USB LZV803 1 1 1 1 2 AC AC DC 1 USB 2 194...
Страница 196: ...NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH 10 30 USB USB USB 196...
Страница 197: ...197...
Страница 198: ...198...
Страница 199: ...199...
Страница 204: ......
Страница 205: ......