System Setup - 9705
K
S
S
- Ma
rch
2
0
06
- Ed
itio
n 1
Operator Manual
HIAC
Note:
Sizes entered must be up to two decimal places between 1.3 and 200.0 or an error message
will be invoked.
4) Use the backspace key to delete any unwanted text in the field. Enter the particle size
(in microns) to be measured, up to two decimal places (such as 1.55).
5) Press
Enter
to save changes and close the
Particle Size Entry
screen.
6) Press the
Return
icon to return to the main screen.
3.5.2
Editing a Size
1) From the Main Screen, select the
Setup
icon
> Sizes
icon to invoke the
Size Settings
screen, shown in
2) Touch the
Edit Size
icon from the
Sizes
tab to invoke the
Particle Size Entry
screen.
3) Use the touch screen to edit a size.
Note:
Sizes entered must be up to two decimal places between 1.3 and 200.0 or an error message
will be invoked.
4) Use the backspace key to delete any unwanted text in the field. Enter the particle size
(in microns) to be measured, up to two decimal places (such as 1.55).
5) Press
Enter
to save changes and close the
Particle Size Entry
screen.
6) Press the
Return
icon to return to the main screen.
3.5.3
Deleting a Size
1) From the Main Screen, select the
Setup
icon
> Sizes
icon to invoke the
Size Settings
screen, shown in
2) Highlight a size from the list on the left by touching it.
3) Press
Delete
Size
to delete the size. The size disappears from the list at left.
4) Press the
Return
icon to return to the main screen.
3.6
Setting Flow Rate
Note:
The 9705 is not flow rate independent. It has been calibrated to a specific flow rate. To operate
the 9705 at a different flow rate, calibrate the 9705 to that specific flow rate. PharmSpec will
store multiple calibration curves for this purpose.
CAUTION:
If the flow rate is set to a value other than the rate corresponding to the calibration curve, the
system response may be inaccurate.
1) From the Main Screen, select the
Setup icon
> Syringe
icon to invoke the
Syringe
Settings
2) Touch the
Nominal Flow Rate
field to invoke a
Flow Rate Entry
keypad screen.
Содержание HIAC 9705
Страница 1: ...HIAC 9705 Liquid Particle Counting System Operator Manual...
Страница 2: ...9705 KSS March 2006 Edition 1 Operator Manual...
Страница 6: ...6 of 82 Table of Contents 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 10: ...10 of 82 Manual Overview 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 22: ...22 of 82 Initialization 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 60: ...60 of 82 Operating the 9705 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 64: ...64 of 82 Maintenance Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 72: ...72 of 82 Service Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 78: ......
Страница 81: ...9705 User Notes 81 of 82 KSS March 2006 Edition 1 HIAC Operator Manual User Notes...