Hach H128 Скачать руководство пользователя страница 1

Continue as follows for two-point calibration

TIP:

When using two-point calibration, always bracket the expected pH value of the sample.  

For example, if the expected pH of a sample is pH 5.00, use pH 7.00 and pH 4.01 buffer.

8.

Apply one drop of either pH 4.01 or pH 10.01 buffer to the probe tip.

9.

Depress the CAL 2 button with the tip of the clip. The meter will automatically identify the 
buffer value.

10.

The CAL icon will flash until a stable reading is obtained.  When the CAL disappears, 

calibration of the second point is complete.

11.

Rinse the probe in tap water and blot dry. To measure the pH of a sample, go to Step 12. 
To continue with a three point calibration, go to Step 13.

12. 

Now measure the sample pH by placing a single drop on the probe tip or dipping the probe tip 
into solution.

Continue as follows for three point calibration

TIP:

Use three-point calibration when the expected sample pH values are completely unknown or 

range both above and below pH 7.00.

13.

Apply one drop of the remaining buffer (either pH 4.01 or pH 10.01) to the probe tip.

14.

Depress the CAL 2 button with the tip of the clip. The meter will automatically identify the 
buffer value.

15.

The CAL icon will flash until a stable reading is obtained.  When the CAL disappears, calibration 
of the third point is complete.

16.

Rinse the probe in tap water and blot dry. 

17.

Now measure the sample pH by placing a single drop on the probe tip or dipping the probe tip 
into solution.

Model H128 

pH Meter with Silicon Chip Sensor

IQ Scientific Instruments

Calibration and Measurement Instructions

Parts

When finished, always rinse the probe and replace the protective cap.

C

Remove cap from probe
tip and turn on the meter.

Rinse the probe in tap
water and blot dry.

Apply one drop of 7.00
buffer to the probe tip.  Be
sure to cover both the
reference junction and the
pH sensor.

Depress the CAL 1 button
with the tip of the clip.

The CAL and 7.00 will
flash until a stable reading
is obtained.   When the
CAL disappears and 7.00
stops blinking, calibration
is complete. 

Rinse the probe in tap
water and blot dry.
To measure the pH of a
sample, go to Step 7. 
To continue with a two
point calibration, go to
Step 8.

Now measure the sample
pH by placing a single
drop on the probe tip or
dipping the probe tip
into solution.   

1

2

3

4

5

6

7

6. Protective Cap

7. Clip

8. Bubble

9. POWER Button

10. CAL 1 Button

1.

Probe Tip

pH sensor and reference electrode.

2.

pH Sensor

Fast response silicon chip pH sensor with built-in temperature sensor.

3.

Reference Electrode

Replaceable saturated KCl Ag/AgCl electrode.

4.

Reference Junction

Measurement is made when the reference junction and the pH sensor are
immersed in the solution.

5.

LCD Display

CAL 

(blinking): Calibration in process.  Measurement cannot be made.

CAL

(off):  Measurement can be made.

BAT

:  Low battery indicator.  Replace batteries.

6.

Protective Cap

Always replace the cap after use.

7.

Clip

Use the tip of the clip to depress the CAL button.

8.

Bubble

The reference solution bubble indicates the life of the reference electrode.

9.

POWER Button

Press to turn the meter on or off.

10. 

CAL 1 Button

Button for pH 7.00 calibration.

11.   CAL 2 Button

Button for pH 4.01 or pH 10.01 calibration.

11. CAL 2 Button

1. Probe Tip

2. pH Sensor

3. Reference Electrode

4. Reference Junction

5. LCD Display

OPERATING INSTRUCTIONS

pH Sensor

Reference Junction

5600 Lindbergh Dr., Loveland, CO  80539
T

Te

ecch

hn

niicca

all  S

Su

up

pp

po

orrtt::

(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
[email protected] • www.hach.com

International customers:
Tel +001 (970) 669-3050
Fax +001 (970) 669-2932

[email protected]

DOC022.53.80061

mini_man_H128  9/8/09  9:29 AM  Page 1

Отзывы: