39
Invalid Calibration Indication
During a calibration, if the offset is greater than ±300 mV, or if the slope is outside
the range of 44.2 to 89.5 mV, an invalid message is displayed and the calibration is
not accepted.
2.3.1.1
Two-Point Calibration with Auto Recognition of Hach Buffers
The buffers used in a two point calibration should bracket the pH range of the sample.
The following procedure uses the pH 4.01 (Cat. No. 22269-66) and pH 7.00
(Cat. No. 22270-66) Hach buffers supplied with the instrument, but a pH 10.00 Hach
buffer (Cat. No. 22271-66) can be used as one of the buffers. If buffers other than the ones
listed above are used, refer to Section 2.3.1.3 Calibration Procedure Variations.
1. Prepare a pH 4.01 Hach Buffer (Cat. No. 22269-66) and a pH 7.00 Hach Buffer
(Cat. No. 22270-66) by dissolving the contents of one powder pillow in 50 mL of
deionized water, and mix.
2. Remove the electrode from the sample solution and clean if needed (see Section 8.1
3. Rinse the electrode with deionized water and then place the electrode in prepared pH
4.01 buffer (Cat. No. 22269-66).
4. Press
CAL
to call up the main calibration menu.
The display shows:
5. Press
ENTER
. The display shows:
6. Press
ENTER
. The display will show the current pH value.
7. After the electrode stabilizes (refer to the stability indicator note), press
ENTER
to
accept this calibration point.
8. Rinse the electrode with deionized water then place the electrode in prepared pH 7.00
buffer solution (Cat. No. 22270-66).
9. Press
ENTER
. The display shows:
10. Press
ENTER
. The display will show the current pH value.
11. After the electrode stabilizes (refer to the stability indicator note), press
ENTER
to
accept this calibration point.
12. Press
pH/mV
to return the meter to normal operations.
13. Rinse the electrode with deionized water and return it to the sample.
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......