35
Press
ENTER
.
The DATABIT menu is 7 or 8. Select the value by pressing the
DOWN ARROW
key or the
UP ARROW
key.
Press
ENTER
.
The STOPBIT menu is 1 or 2. Select the value by pressing the
DOWN ARROW
key or the
UP ARROW
key.
Press
ENTER
.
Press
ENTER
followed by the
DOWN ARROW
key or the
UP ARROW
key to toggle between
ENABLED and DISABLED. Press
ENTER.
The PRinT INTerval setting range is 00 to 99 minutes. 00 turns the print interval off. Make
a selection by pressing the
RIGHT ARROW
key followed by the
DOWN ARROW
key or the
UP ARROW
key.
Press
ENTER
.
PRinT LiSTing appears on the display.
Press
ENTER
. A listing of the instrument setup is sent to the printer. Then, OK appears on
the display momentarily and the display automatically backs out to BAUD.
Press
CLEAR
followed by the
DOWN ARROW
key. This backs out to:
Press
ENTER
followed by the
DOWN ARROW
key or the
UP ARROW
key to toggle between
CONTROL and RECORD2 modes.
Press
ENTER
.
Press
ENTER
.
This displays the pH SIGnal AVeraGe value (00 to 10 seconds), 03 seconds is the default.
Change the display by pressing the
RIGHT ARROW
key followed by the
DOWN ARROW
key
or the
UP ARROW
key.
Press
ENTER
followed by
CLEAR
to back out to the top level menu, or press the
pH/mV
key
to exit the menu.
2.2.6
Keylock
Note: The initial keyboard
lockout code is 0000, but is
user programmable. Refer
to Section 9.1.1 Test
Diagnostic Menu for
information on
programming the keyboard
lockout code.
To prevent unauthorized or accidental programming changes, a keyboard lockout or secure
mode is provided using a password. In the locked mode, all program menus may be
viewed, but no changes to the settings or operations are allowed. If a change is attempted,
the display will flash LOCKED to indicate that the keyboard is locked out.
The keyboard may be locked, unlocked, or temporarily (30 minutes) unlocked.
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......