31
Press
ENTER
.
The current value for SET PoiNT appears on the display. Change the value by pressing the
RIGHT ARROW
key and the
DOWN ARROW
key or the
UP ARROW
key.
Press
ENTER
to save the SET PoiNT value and advance to:
Press
ENTER
.
The current value for DEAD BanD appears on the display. Change the value by pressing
the
RIGHT
ARROW
key and the
DOWN
ARROW
key or the
UP
ARROW
key.
Press
ENTER
to save the DEAD BanD value and advance to:
Press
ENTER
.
The current setting for the Controller PRESET output appears on the display. Change
the value by pressing the
RIGHT ARROW
key and the
DOWN ARROW
key or the
UP ARROW
key.
Press
ENTER
to save the PRESET value and advance to:
Press
ENTER
.
The current setting for the Controller GAIN appears on the display. Change the value by
pressing the
RIGHT ARROW
key and the
DOWN
ARROW
key or the
UP ARROW
key.
Press
ENTER
to save the GAIN value and advance to:
Press
ENTER
.
The current value for the Controller RESET action appears on the display. Change
the value by pressing the
RIGHT ARROW
key and the
DOWN
ARROW
key or the
UP ARROW
key.
Press
ENTER
to save the RESET value and advance to:
Press
ENTER
.
The current value for the Controller RATE action appears on the display. Press the
RIGHT
ARROW
key and the
DOWN ARROW
key or the
UP
ARROW
key to change the value.
Press
ENTER
to save the RATE value and advance to:
Press
ENTER
.
Toggle between the TITRation CURve mode selections (, linear -, curve) by
pressing the
DOWN ARROW
key or the
UP ARROW
key.
The user must choose or linear - unless a valid set of points is entered. Press
ENTER
to save the value and advance to:
Press
ENTER
.
Select the number of points desired on the TITRation CURve by pressing the
RIGHT
ARROW
key and the
DOWN ARROW
key or the
UP ARROW
key.
(Range: 3 to 20)
Note: The following data
entry sequence is
dependent upon the
number of points selected
above (9).
Press
ENTER
. The display asks for percent of reagent addition (X axis) and pH or mV
value points (Y axis). Percentages and values must be entered in ascending and
descending order with a new entry for each point selected. The controller does not accept
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......