![Hach DR 6000 Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875080.webp)
80
Expanded programs
6.1.1
Program a user method
All input steps and their significance and options are explained in
the following sections.
1.
Select
New
in the "Program Options" menu.
Program Number:
The program number is a specific test number, with which the
program can subsequently be called up from the selection list in
the
User Programs
menu or the
Favorites
menu.
2.
Use the alphanumeric keypad to enter a program number
between 9000 and 9099. The lowest available number
appears automatically.
3.
Press
OK
.
Note: Should a program number already be assigned to another
user program, the query occurs as to whether the existing program
should be replaced. If you tap on OK, the existing program is
overwritten.
Program Name:
4.
Enter a program name via the alphanumeric keypad. The
name can be a maximum of 28 characters long.
5.
Press
Back
to go back to the previous program point or
press
Next
to continue with the input of the program data.
Program type:
6.
Select the required program type (
Table 8
) and press
Next
.
7.
If the options
Single Wavelength
(
or
Multi-Wavelength
(
) are
selected, define the programming parameters:
•
Unit
•
Wavelength
•
Calculation formula
•
Wavelength
x
•
Concentration factor Kx
•
Concentration solution
•
Chemical form
•
Calibration equation
For more information on free programming parameters refer to
.
Содержание DR 6000
Страница 2: ......
Страница 16: ...16 General information ...
Страница 132: ...132 Expanded programs ...
Страница 138: ...138 Maintenance 7 4 Replace fuses 1 Remove the cover below the power switch containing the fuses ...
Страница 142: ...142 Troubleshooting ...
Страница 145: ...145 Section 10 Contact ...
Страница 146: ...146 Contact ...
Страница 152: ...152 Help Guide ...
Страница 156: ...156 Index ...