Standard programs
5.2.8
Power Management
1.
Press
Power Management
in "Instrument Setup".
2.
Select
Sleep Timer
.
3.
Select the required time span after which the instrument will
go into energy-saving sleep mode if it is not used (refer to
), and confirm with
OK
.
4.
Select
UV Lamp Save
.
5.
Select the required time span after which the UV lamp will be
powered off if it is not used, and confirm with
OK
.
Note: To protect the UV lamp, do not switch it on and off frequently.
The UV lamp requires time to warm up.
6.
Select
Visible Lamp Save
.
7.
Select the required time span after which the VIS lamp will
be powered off if it is not used, and confirm with
OK
.
Note: The manufacturer recommends the "Automatic" setting for
this. This means the VIS lamp is only switched on momentarily to
perform the measurement.
5.2.9
Lamp control
With the lamp control the UV lamp and the VIS lamp can be
switched on or off manually.
1.
Tap
Lamp Control
in the "Instrument Setup".
2.
Select
VIS-Lamp
or
UV-Lamp
.