Expanded programs
the wavelength. The data associated with each data point can
also be shown as a table.
6.5.1
Set up the wavelength scan
Press
Wavelength Scan
in the Main Menu. Press
Options
for
Parameter Setup.
Table 14 Options during wavelength scan
Option
Description
More
Further options are displayed
Folder icon
To store the scan data
Reference Off/On
From the displayed list of stored scans, a record is selected for use as a reference
scan/superimposed scan. This can be highlighted or shown in the background in comparison
with the actual measured scan.
Note: This option is only available when there are stored scans with the same wavelength range
and step.
Enter the wavelength spectrum and scan interval
Timer icon
This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.
View
Table/Graph
Enables the switching between the display of tabular scan data (wavelength/absorbance) and
the graphical representation of the curve.
Note: "View Table" is activated after the first measurement.
Cursor Mode
To select
Track
or
Peak/Valley
. The selection for this menu option determines to which points
on the graph the cursor moves.
Send Data
To send data to a printer, computer or USB memory stick (Type A)
Integral: On/Off
The integral gives the area and the derivative of the integral gives the original function
Scale & Units
Scale:
In the automatic scaling mode, the y-axis is automatically adapted so that the total scan
is shown.
The manual Scaling mode allows sections of the scan to be displayed.
Units:
Choice of absorbance or transmittance.
Reading Mode
Single Reading
: One measurement result is shown after tapping on
Read
.
Carousel 1 inch square
: Optional measurement of carousel insert with up to 5 square cells.
Carousel 1 cm square
: Optional measurement of carousel insert with up to 7 square cells.
Recall measurement
data
Recall saved measurement data, wave-length scans or time courses, see
.
Instrument Setup mode
Basic data of the instrument, refer to
Содержание DR 6000
Страница 2: ......
Страница 16: ...16 General information ...
Страница 132: ...132 Expanded programs ...
Страница 138: ...138 Maintenance 7 4 Replace fuses 1 Remove the cover below the power switch containing the fuses ...
Страница 142: ...142 Troubleshooting ...
Страница 145: ...145 Section 10 Contact ...
Страница 146: ...146 Contact ...
Страница 152: ...152 Help Guide ...
Страница 156: ...156 Index ...