Reagent preparation

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.

Prepare the arsenic absorber solution as follows:

1.

Weigh 1.00 g of silver diethyldithiocarbamate on an analytical balance.

2.

Add the powder to a 200-mL volumetric flask. Dilute to the mark with pyridine.

3.

Mix well to dissolve. Keep the reagent, tightly sealed, for a maximum of 1 month in an
amber bottle.

Calibration

This method requires a user-prepared calibration curve. Enter the calibration curve into
the instrument as a user program. Make a new calibration curve when a new lot of
reagents is used.

Prepare the standard solutions

Prepare the standard solutions for calibration as follows.
Items to collect:

1000-mg/L Arsenic Standard Solution

1000-mL volumetric flask

500-mL volumetric flasks (3), Class A

1–10 mL TenSette pipet and tips

Deionized water

1.

Prepare a 10.0-mg/L Arsenic Standard Solution as follows:

a.

Use a pipet to add 10.0 mL of the 1000-mg/L Arsenic Standard Solution into a
1000-mL volumetric flask.

b.

Dilute to the mark with deionized water.

2.

Prepare three calibration standard solutions (0.02, 0.04 and 0.20 mg/L arsenic) as
follows:

a.

Use a pipet to add 1.0 , 2.0 and 10.0 mL of the 10.0-mg/L Arsenic Standard
Solution into three different 500-mL volumetric flasks. Use Class A glassware.

b.

Dilute each flask to the mark with deionized water. Mix well.

Enter the calibration as a user program

After the calibration standards are prepared, make a user program to store the calibration
information in the instrument. Select the user program to measure the concentration of
samples. The steps that follow are general instructions for all instruments. Refer to the
user manual for the instrument that is used for the correct menu options.

1.

Use the test procedure to prepare the standard solutions for measurement. Make
sure to distill the standard solutions as shown in the test procedure.

2.

Go to 

User Programs

.

3.

For the initial calibration, make a new user program. Set up the basic information for
the new program:

Option

Description

Program number

Enter an available number for the user program.

Arsenic Silver Diethyldithiocarbamate Method (0.200 mg/L)

5

Содержание DR 6000

Страница 1: ...Before starting A manual calibration is necessary for this procedure refer to Calibration on page 5 Make a new calibration and update the user program for each new lot of arsenic absorber solution Pre...

Страница 2: ...p the preserved samples at room temperature for a maximum of 6 months Before analysis adjust the pH to approximately 7 with 5 N sodium hydroxide solution Correct the test result for the dilution cause...

Страница 3: ...ric Acid ACS to the distillation flask 10 Use a serological pipet to add 1 mL of Stannous Chloride Solution to the distillation flask 11 Use a serological pipet to add 3 mL of Potassium Iodide Solutio...

Страница 4: ...n the blank sample cell 22 Insert the blank into the cell holder Zero 23 Push ZERO The display shows the non zero intercept that is calculated from the user entered calibration curve 24 Prepare the sa...

Страница 5: ...s follows a Use a pipet to add 10 0 mL of the 1000 mg L Arsenic Standard Solution into a 1000 mL volumetric flask b Dilute to the mark with deionized water 2 Prepare three calibration standard solutio...

Страница 6: ...FORCE ZERO is off 12 If the graph is acceptable select DONE EXIT 13 Save store the user program When the user program is selected in the test procedure the calibration curve is used to measure the sa...

Страница 7: ...s stopper 1 each 1457453 Flask volumetric Class A 200 mL 1 each 1457445 Flask volumetric Class A 500 mL 6 each 1457449 Pipet filler safety bulb 1 each 1465100 Pipet serological 5 mL 1 each 53237 Pipet...

Страница 8: ...ANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll...