the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference
to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following
techniques can be used to reduce interference problems:

1.

Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.

2.

If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different
outlet.

3.

Move the equipment away from the device receiving the interference.

4.

Reposition the receiving antenna for the device receiving the
interference.

5.

Try combinations of the above.

2.2 Product overview

Respirometric Biological Oxygen Demand (BOD) is a test that measures
the quantity of oxygen consumed by bacteria that oxidize organic matter
in a water sample. The test is used to measure waste loadings at
wastewater treatment plants and to examine the efficiency of wastewater
treatment.
The instrument is sealed to prevent external atmospheric pressure
changes in the test bottle. The pressure in the sample bottles is
monitored. Bacteria in the sample use oxygen when they consume
organic matter. This oxygen consumption causes the pressure in the
bottle head space to drop. The pressure drop correlates directly to BOD.
During a test period, stir bars mix the sample and cause oxygen to move
from the air in the bottle to the sample. This helps simulate natural
conditions.
Carbon dioxide is a result of the oxidation process and can interfere with
a measurement. The instrument continuously removes carbon dioxide
from the system so that the monitored pressure difference stays
proportional to the quantity of oxygen used. Pressure changes in the

closed system are shown graphically in milligrams per liter (mg/L) on a
liquid crystal display. The instrument gives 360 uniform data points over
the selected time period.
The instrument adjusts for any negative errors produced when heat is
applied to a sample. The instrument does not start the test until the
temperature gets to equilibrium.

2.3 Product components

Make sure that all components have been received. If any of these items
are missing or damaged, contact the manufacturer or a sales
representative immediately.

• BODTrak

 II instrument

• A UL/CSA approved 115 VAC power cord with a NEMA 5-15P style

plug

• A 230 VAC harmonized power cord with a continental European plug
• Power supply, auto-switching between 115 V and 230 V
• Seal cups (6x)
• BODTrak II amber sample bottles (6x)
• BODTrak II magnetic stir bars (6x)
• Spatula scoop
• Nutrient buffer solution pillows (1 pkg)
• Potassium hydroxide pellets (1 container)

Section 3 Installation

3.1 External connections

Figure 1

 shows the locations of the power switch and external

connections.

English

   

5

Содержание BODTrak II

Страница 1: ...DOC022 53 80363 BODTrak II 11 2020 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...tions 5 3 2 Connect the RS232 interface 6 3 3 Bottle connections 6 Section 4 User interface 6 4 1 Channel selection keys 7 4 2 Arrow keys 7 4 3 ON key 7 4 4 OFF key 7 Section 5 Startup 7 5 1 Turn the...

Страница 4: ...10 4 Toxic materials 16 10 5 Chlorine 16 10 6 pH effect 16 10 7 Supersaturation 16 Section 11 Maintenance 17 11 1 Cleaning 17 11 2 Storage 17 Section 12 Troubleshooting 17 12 1 Incorrect BOD curves 17...

Страница 5: ...rd 150 mg L each of glucose and glutamic acid Number of samples 44 Number of analysts 1 Number of BodTrak II instruments 6 Results Mean of 235 mg L BOD Distribution 11 mg L or range of 224 to 246 mg L...

Страница 6: ...references the instruction manual for operation and or safety information Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems Return ol...

Страница 7: ...es is monitored Bacteria in the sample use oxygen when they consume organic matter This oxygen consumption causes the pressure in the bottle head space to drop The pressure drop correlates directly to...

Страница 8: ...s if the device cannot continuously receive at 9600 baud Note The use of the specified cable or an equivalent shielded cable is necessary to meet radio frequency emissions requirements 3 3 Bottle conn...

Страница 9: ...test length and range 4 3 ON key To go to the range selection menu push ON from a channel display screen Then push and hold ON to start the test for the selected channel 4 4 OFF key When a test is in...

Страница 10: ...zard select Delimited as the file type the start row in the spreadsheet and Windows ANSI as the file origin Click NEXT 4 Click the check boxes for Space delimiter and Treat consecutive delimiters as o...

Страница 11: ...propriate to the chemicals that are handled Refer to the current safety data sheets MSDS SDS for safety protocols C A U T I O N Chemical exposure hazard Dispose of chemicals and wastes in accordance w...

Страница 12: ...level the 0 to 35 mg L BOD range is decreased to 0 to 25 mg L BOD Adjustment is not necessary for other test ranges Refer to Special considerations on page 15 for special considerations including samp...

Страница 13: ...e size If samples are seeded prepare a seed blank before a sample is prepared Treat the seed blank the same as any other sample and omit step 5 Use Hach BOD Standard Solution Ampules for Manometric Me...

Страница 14: ...eded use a tensette pipet to add the correct quantity of seed to the graduated cylinder Table 5 6 If necessary add more nutrient or buffer Do not add more than a total volume of 50 mL seed nutrient bu...

Страница 15: ...s are rotating If not lift the bottle up then set it down again 7 Push and hold the LEFT and RIGHT arrows at the same time to go to the instrument setup menu Note Set the time and date if necessary Se...

Страница 16: ...s and the seed blank bottle and GGA bottle when applicable The results are shown Note Treat the seed blank the same as all other samples If the sample was pre diluted apply a dilution factor to the in...

Страница 17: ...r to the expanded version of the manual for information about specific procedures Figure 3 shows typical curves through a 10 day test period For incorrect curves refer to Figure 4 on page 18 Figure 3...

Страница 18: ...4 Toxic materials Industrial and chlorinated samples often contain toxic substances and require special considerations when running BOD tests Toxic materials in the sample will cause decreased BOD va...

Страница 19: ...onized or distilled water Clean the sample bottles Clean the sample bottles and caps with a brush water and a mild detergent Flush the sample bottles with fresh water followed by a distilled water rin...

Страница 20: ...fluent is in the 0 70 mg L range while influent is in the 0 700 mg L range when the BOD of the sample is more than 700 mg L prepare a sample dilution Refer to the Sample dilution section in the expand...

Страница 21: ...leak makes the system unresponsive If such a condition occurs examine the seal cup and bottle cap for contamination or damage Section 13 Replacement parts and accessories 13 1 Replacement parts Descr...

Страница 22: ...size 2 1 68700 Buret straight stopcock 25 mL 1 1405940 Clamp buret double 1 32800 Cylinder graduated 10 mL 1 50838 Cylinder graduated 25 mL 1 50840 Cylinder graduated 50 mL 1 50841 Cylinder graduated...