Page 99
8.1.4.2
Send Fault Archive
This menu is used to download the fault archive. The communication port parameters used are those set up
in the Data Program menu. All items in the fault archive will be downloaded unless the user interrupts the
downloading with the Pause Sending or Stop Sending functions.
1. Start Sending.
Press the ENTER key to start downloading the data.
2. Pause Sending.
Press the ENTER key to interrupt the downloading of data. Press again to continue
downloading. If the downloading is interrupted for more than 60 seconds, then the downloading is
automatically resumed.
3. Stop Sending.
Press the ENTER key to stop downloading the data.
OUTPUT ITEMS
is the number of events currently downloaded. The maximum number of events in the fault
archive is 99.
When external MMC/SD flash memory card is being used as the output device, the fault archive will be saved
into the card in text format and named as “FARCH.TXT” by default.
8.1.4.3
Send Configuration
This menu is used to download the configuration data in the BioTector. The communication port parameters
used are those set up in the Data Program menu. All the configuration data will be downloaded unless the
user interrupts the downloading with the Pause Sending or Stop Sending functions.
1. Start Sending.
Press the ENTER key to start downloading the data.
2. Pause Sending.
Press the ENTER key to interrupt the downloading of data. Press again to continue
downloading.
Note:
If the downloading is interrupted for more than 60 seconds, then the downloading is automatically
resumed.
3. Stop Sending.
Press the ENTER key to stop downloading the data.
OUTPUT ITEMS
is the number of events currently downloaded.
When external MMC/SD flash memory card is being used as the output device, the system configuration will
be saved into the card in text format and named as “CNFG.TXT” by default.
8.1.4.4
Send All Data
This menu is used to download all data (i.e. Reaction Archive, Fault Archive, Configuration and System
Diagnostics) in one simple operation. The communication port parameters used are those set up in the Data
Program menu. All the diagnostic data will be downloaded unless the user interrupts the downloading with the
Pause Sending or Stop Sending functions.
Unlike individual data download (e.g. reaction archive, fault archive and configuration), which are downloaded
in the selected system language, all data is downloaded in English language only.
1. Start Sending.
Press the ENTER key to start downloading the data.
2. Pause Sending.
Press the ENTER key to interrupt the downloading of data. Press again to continue
downloading. If the downloading is interrupted for more than 60 seconds, then the downloading is
automatically resumed.
3. Stop Sending.
Press the ENTER key to stop downloading the data.
OUTPUT ITEMS
is the number of events currently downloaded.
When external MMC/SD flash memory card is being used as the output device, the all data will be saved into
the card in text format and named as “ALLDAT.TXT” by default.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...