Page 96
8.1.4.1
Send Reaction Archive
S E N D R E A C T I O N A R C H I V E 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < S T A R T D A T E 0 1 - 0 9 - 0 2
2 N U M B E R O F E V E N T S 1 2 3
3 S T A R T S E N D I N G
4 P A U S E S E N D I N G
5 * S T O P S E N D I N G
O U T P U T # 1 2 3 I T E M S
This menu is used to download the reaction archive. The communication port parameters used are those set
up in the Data Program menu.
1. Start Date.
This is the start date of the first item to be downloaded. The default date is the current date on
the BioTector, which can be changed by the user. The newest event is downloaded first when
downloading data.
2. Number of Events.
This is the number of events to be downloaded. The default is the number of events
in the reaction archive, which can be edited by the user.
3. Start Sending.
Press the ENTER key to start downloading the data.
4. Pause Sending.
Press the ENTER key to interrupt the downloading of data. Press again to continue
downloading. If the downloading is interrupted for more than 60 seconds, then the downloading is
automatically resumed.
5. Stop Sending.
Press the ENTER key to stop downloading the data.
OUTPUT ITEMS
is the number of events currently downloaded. The maximum amount of events is 9999.
When external MMC/SD memory card is being used as the output device, the reaction archive will be saved
int
o the card in text format and named as “RARCH.TXT” by default.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...