Page 48
4.2.4.3
VOC (POC) Analysis
BioTector Volatile Organic Carbon (VOC) analysis type is a combination of TC analysis followed by a TIC &
TOC (NPOC) analysis. The VOC result obtained from the VOC analysis type represents the Purgeable
Organic Carbon (POC) content of the sample. When both TC and TIC & TOC analysis are complete, the
flowing data is available:
TC result, as measured and displayed from the TC analysis.
TIC result, as measured and displayed from the TIC & TOC analysis. TOC result obtained from the
TIC & TOC analysis represents the NPOC.
TOC result including the VOC, which is calculated from the difference between the TC and TIC:
TOC
v
= TC
– TIC
The TOC result displayed in VOC analysis type includes the purgeable organic carbon (POC) present
in the sample. In other words the TOC result, obtained in VOC analysis type, is the sum of NPOC and
POC:
TOC
v
= NPOC + POC
VOC (POC) result, which is calculated from the difference between the measured TC (from the TC
analysis), and the sum of measured TIC and measured NPOC (from the TIC & TOC analysis):
VOC (POC) = TC
– (TIC + NPOC)
The NPOC result, as measured from the TIC & TOC analysis, is not displayed, it is only used to
calculate the VOC (POC) element in the sample. All displayed results can be programmed in the
system and sent as 4-20mA output signals to an external device.
BioTector analysis types are optional features. If the BioTector is built as a TIC &
TOC system only, modifications in the system configuration will be required for VOC
analysis to be possible.
BioTectors built with the VOC analysis option can be programmed on site to operate
with the TIC & TOC only or TC only analysis types.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...