Page 132
8.3.10
CO
2
Analyzer
ANALYSIS GRAPH SCALE 10000ppm
ANALYSIS GRAPH SCALE determines the scale of the y-axis
“CO2 analyzer’s ppm-CO2 readings” on the Analysis Graph
Screen (see section
This function allows the system to display CO2 data peaks in
optimal resolution on the LCD screen.
The ANALYSIS GRAPH SCALE is independent of the CO2
ANALYZER RANGE described below. Note that, in certain
circumstances, when the displayed CO2 peaks on screen
exceeds the scale of the graph, the CO2 analyzer keeps
measuring and integrating the CO2 readings to obtain the TOC
result, without losing any CO2 data.
BAUDRATE 9600
BAUDRATE (9600bps by default) specifies the data
communication signaling speed of CO
2
Analyzer with the RS232
communication interface.
CO2 ANALYZER RANGE 10000ppm
CO2 ANALYZER RANGE defines the full scale range of the CO
2
analyzer installed in the BioTector.
CO2 ANALYZER CAL
The CO2 ANALYZER CALIBRATION menu allows the user to
change CO
2
Analyzer range and the CO
2
analyzer zero and
span parameters when necessary. If required, please contact
manufacturer or distributor for details of the procedures.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...