8.1.5 Input/Output Status
Input/Output Status menus are used for monitoring the analog and digital inputs and outputs for advanced
diagnostics purposes.
Digital Input
The Digital Input menu allows the engineer to monitor the system digital inputs. This feature is useful for the
system trouble-shooting or diagnostics. On the screen the digital inputs are organized in columns and rows
with their code, logical states and function. Each input name is composed of the “DI” prefix and two-decimal
index, which identifies the input. For exampl
e, the digital input named “DI06” is digital input 6, which is the
ENTER key. Its logical state is shown as 0, therefore the ENTER key is not pressed or activated. When
ENTER key is pressed, its logical state will be shown as 1.
In the idle or open circuit state, all the system digital inputs are set to logical state 0. The active or closed
circuit state is logical state 1. The programmable digital inputs are marked as [PROGRAMMABLE] in this
menu.
Digital Output
The Digital Output menu allows the engineer to monitor the system digital outputs. This feature is useful for
the system trouble-shooting or diagnostics. On the screen the digital outputs are organized in columns and
rows with their code, logical states and function. Each outpu
t name is composed of the “DO” prefix and two-
decimal index, which identifies the output. For instance
, the digital output named “DO34” is digital output 34,
which is used to control the heater of the Ozone Destructor. Its logical state is shown as 0 and 1, which
stands for OFF and ON respectively. Upon system power up or reset, all system digital outputs are set to
logical state 0.
Analog Input
The Analog Input menu allows the user to monitor the system analog inputs. This feature is useful for the
system trouble-shooting or diagnostics. On the screen the analog inputs are organized in columns and rows.
Each analog input has three parameters. The first is ADC converter digital value, the second is the input
voltage measured in volts and the third is the function. The BioTector uses a 12-bit ADC, therefore the range
of the digitized inputs are 0-4095. The voltage range is 0 to 5.00 Volts.
Analog Output
The Analog Output menu allows the user to monitor the system analog outputs. This feature is useful for the
system trouble-shooting or diagnostics. On the screen, each analog output has three parameters. The first is
DAC converter digital value, the second is output voltage measured in volts and the third is the function. The
BioTector uses a 12-bit DAC, therefore the range of the digitized outputs are 0-4095. The voltage range is 0
to 10.00 Volts.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...