Installation
19
Note:
Primary components may
become hot under high line voltage
and ambient temperature
conditions. ac wiring insulation must
be rated a minimum of 95 °C.
6.
Tighten the connectors using a small flat-head screwdriver. Firmly tug on
each wire to ensure solid connections.
7.
Reinstall the shielding plate and front cover if all of the wiring is complete.
8.
See
for information on setting alarms.
Figure 14
Relays Terminal Strip
4.4.5 Analog and RS485 Output Connections
This transmitter is designed for either a 4–20 or 0–20 mA output with a maximum
800 ohm load. Make output connections with twisted-pair shielded wire. Connect
the shield at the recorder or controlled component end only. On the opposite end,
remove the shield and insulate the remaining wires.
Note:
The RS485 card is factory
installed, it does not need to be
re-installed.
1.
Remotely disconnect all instrument power before opening the analyzer door.
2.
Remove the front cover.
3.
Strip the insulation on the outer jacket back 7 inches. Remove the conductors
from the shield.
4.
Cut the shield to 1 inch. Connect it to the ground on the shielding plate.
Note:
Use a twisted-pair, shielded
cable. Use of non-shielded cable
may result in radio frequency
emission or susceptibility levels
higher than allowed.
5.
Strip the insulation on the remaining wires back ¼-inch.
6.
Route the wire through the front left conduit opening. Use the provided
cord-connected strain reliefs to ensure adequate sealing. Use appropriate
NEMA sealing hardware.
7.
Insert the wire ends into the Analog or RS485 connector until the insulation
seats against the connector. Refer to
and
.
8.
Tighten the connectors using a small flat-head screwdriver. Firmly tug on
each wire to assure solid connections.
9.
Reinstall the shielding plate and front cover if all of the wiring is complete.
IC1
J1
D3
D4
RE4
RE3
D2
RE2
RE1
V1
V2
V3
V4
D1
C8
L2
C6
Z6
Model : 09125-A-2000
Serial No : 0104 549
Ind : D
Wiring located behind the guide plate shall be
rated at 80 C (176 F) or higher.
Vac : 100 - 240 V
K2
K9
K4
K5
IC2
J1
RP1
Ic4
K6
RP2
Ref
Temp +
Temp –
Counter
Work
Aux
9180
AMP
V1.03
P1
Relays (S1– S4)
S1
S2
S3/Sys. Alarm
S4/Timer