Orbisphere 510/511/512 Analyzers - Calibration Menu
65 of 138
Operator Manual
ORBISPHERE
5.8 External Pressure Calibration (optional sensor)
Two calibration methods for the external pressure sensor
can be selected:
• Two point calibration (recommended)
• One point calibration
By default the two point calibration is selected
Fig 5-3:
Two Point Calibration
Two Point Calibration (recommended)
Connect a certified absolute pressure gauge to the
sample line, and use a certified precision barometer.
• The two point calibration starts with barometric
pressure for the lower point. Expose the external
pressure sensor to the atmosphere.
Enter the barometric pressure read on the
barometer in the upper window and
validate P1
.
• Expose the external pressure sensor to line
pressure, making sure it is exposed to the same
pressure as the certified absolute pressure gauge.
Enter the absolute pressure value read on the
certified absolute pressure gauge in the lower box,
and
validate P2
.
Note:
Any pressure can be used for P1 and P2, but for an
accurate calibration P1 and P2 should be as different as
possible.
Fig 5-4:
One Point Calibration
One Point Calibration
Connect a certified absolute pressure gauge to the
sample line.
See “Generic Terms and Definitions” on
• Expose the external pressure sensor to the line
pressure, making sure it is exposed to the same
pressure as the certified absolute pressure gauge.
Enter the absolute pressure value read on the
certified absolute pressure gauge in lower box, and
validate P1
.
Note:
Any pressure can be used for P1, but for one point it
should be as close as possible to the sample pressure
Содержание ORBISPHERE 510
Страница 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Страница 2: ......
Страница 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 43: ...Orbisphere 510 511 512 Analyzers View Menu 37 of 138 Operator Manual ORBISPHERE 3 View Menu Fig 3 1 View Menu ...
Страница 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 114: ...108 of 138 Global Configuration Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 116: ...110 of 138 Services Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE Fig 11 2 Services Menu Part 2 ...
Страница 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Страница 144: ......