![HACH LANGE H-Series Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/hach-lange/h-series/h-series_basic-user-manual_577212014.webp)
Option
Function
Description
CLr ALL
CAL
Clear calibration
points
Erases all the calibration points. Make sure to calibrate the meter.
1
int
Data log interval
Keeps up to 999 data points in intervals from 1 to 1999 seconds (default =
10) in the data log.
year
Year
Use the number keys to enter the correct year.
date
Date format
Sets the date to mm/dd/yy or dd/mm/yy format.
m/d date
Month and day
Use the number keys to set the correct month and day.
time
Time
Sets the correct time in a 24-hour format. Use the number keys to set the
correct time.
oFF
Automatic
shutdown
Sets the shutdown parameters from 1 minute to 000 minutes (continuous
power). The meter beeps 1 minute before shutdown. Make sure to power
cycle for the automatic shutdown to occur. Automatic shutdown is disabled
during: data logging, Bluetooth
™
transfers and when connected to the USB
port.
Snd
Sound options
Sets sound alerts to on or off. Three sounds tell the user about different
functions:
•
One beep
: Stabilization occurred with the stabilization lock on.
•
Two beeps
: An error occurred. The error code shows on the display.
on page 17.
•
Three beeps
: Measurement stability in calibration mode, regardless of
the stabilization lock setting.
ºC ºF
Temperature
display
Sets the temperature unit from ºC to ºF.
READY
Stabilization lock
Sets the stabilization lock to on or off:
• The "READY" icon shows when an endpoint occurs and the stabilization
lock is set to on.
• After stabilization, the display locks the value and the meter ignores slight
measurement changes in pH, conductivity or TDS.
• Sets the stabilization lock to off during titrations or slight change
detection. The display automatically unlocks after a significant
measurement change is found.
1
This does not erase calibration data for an ISFET probe.
4.
Push
ENTER
to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push CANCEL.
Operation
Configure the Bluetooth
™
wireless connection
Transfer data and control several instrument functions from a remote location to a PC with the
Bluetooth
™
wireless connection.
Note: The arrow icon shows on the display when the meter is in communication with the SmartLogger II software
on a PC. The arrow icon flashes when data is transferred. Refer to the SmartLogger II documentation to setup the
wireless connection to a PC.
1.
Set the power to on. The meter looks for the Bluetooth
™
connection with a PC.
2.
On the PC, select the option to find or add a Bluetooth
™
device. Next, the PC prompts if the user
wants to pair with a found device.
3.
When the PC prompts for a pass key or PIN, enter the PIN (default =
1234
). The PC shows if the
pair is successful and the meter shows the Bluetooth
™
icon.
14
English
Содержание H-Series
Страница 2: ...English 3 Deutsch 20 Français 40 Čeština 59 Dansk 76 Polski 94 български 113 2 ...
Страница 134: ......
Страница 135: ......