DOC023.52.90161

FP 360 sc

User Manual

01/2010, Edition 1B

© HACH-LANGE GmbH, 2010. All rights reserved. Printed in Germany

Содержание FP 360 sc

Страница 1: ...DOC023 52 90161 FP 360 sc User Manual 01 2010 Edition 1B HACH LANGE GmbH 2010 All rights reserved Printed in Germany ...

Страница 2: ......

Страница 3: ...llation with flow cell 15 Section 4 Operation 17 4 1 User interface and navigation 17 4 2 Sensor setup 17 4 3 Sensor data logger 17 4 4 Menu structure 18 4 4 1 SENSOR STATUS 18 4 4 2 SENSOR SETUP 18 4 5 Calibration 20 4 5 1 Factory calibration 20 4 5 2 Process calibration adjustment 20 4 5 2 1 Determination of factors and adjustment of slope 20 4 5 2 2 Multi point calibration 21 4 5 3 Check the ze...

Страница 4: ...es 25 6 2 Warnings 25 6 3 Replacement parts 25 Section 7 Replacement parts and accessories 27 7 1 Sensor options 27 7 2 Replacement parts 27 7 3 Accessories 27 Section 8 Contact information 29 Section 9 Warranty and Liability 31 Appendix A Modbus register 33 Index 35 ...

Страница 5: ...scence calibration standard custom adaptation possible Sensor software Software version From 1 14 Equipment properties Weight Measurement sensor Stainless steel version 2 8 kg titanium version 1 8 kg Flow cell Approximately 0 6 kg including installation board approximately 2 0 kg Pressure range Sensor Max 30 bar Flow cell Max 1 bar Dimensions Measurement sensor Ø length 68 mm x 306 mm 2 68 in x 12...

Страница 6: ... 13 to 131 F Sensor distance wall ground min 100 mm 3 94 in recommended Miscellaneous Cable length 1 5 or 10 m extension cable up to total maximum length of 40 m Connection information Sensor side 8 pin type of protection IP68 PUR Controller side M12 type of protection IP67 Inspection interval Every 2 years 1 year service agreement by request with warranty extension up to 5 years Maintenance requi...

Страница 7: ... could result in death or serious injury C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury N O T I C E Indicates a situation which if not avoided may cause damage to the instrument Information that requires special emphasis This is the safety alert symbol Obey all safety messages that follow this symbol to avoid potential injury If on the instrum...

Страница 8: ...products and are a very specific indicator of oil contamination in water bodies and process water 2 4 Product components Each sensor is supplied with a connector cable with safety sleeve a shackle a basic user manual and a CD The sensor is available in a variety of types refer to section 7 1 on page 27 Refer to Figure 2 to make sure that all components have been received If any of these items are ...

Страница 9: ...l 8 Shackle 2 5 Function test After the components are removed from the package do a function test 1 Connect the connector cable to the sensor 8 pin polarized connector and an appropriate sc controller 5 pin polarized connector refer to section 3 1 on page 11 2 Apply power to the sc controller The display is activated and the sensor goes to measurement mode The sensor ticks quietly and regularly C...

Страница 10: ...do not use recycled paper 4 Vary the distance between the measurement window and the paper The measurement value on the display will change accordingly Note In air the measurement value displayed is not exactly zero due to reflections on the window surface refer to section 4 5 3 on page 21 ...

Страница 11: ...the keyed quick connect fitting Hand tighten refer to Figure 4 Note Connector cables are available in various lengths refer to section 7 2 on page 27 Maximum overall cable length is 40 m 131 23 ft DANGER Personal injury hazard Only qualified personnel should conduct the tasks described in this section of the manual N O T I C E If the sensor is not fully inserted sun protection is recommended in hi...

Страница 12: ...ct the sensor to the controller Figure 5 Pin configuration 1 12 VDC brown 2 Ground black 3 Data blue 4 Data white 5 Screen Screen gray 6 Guide 12 Installation Number Description Standard cable cable color 6 1 2 5 3 4 ...

Страница 13: ...llation of chain mount kit for sensors with cleaning unit 3 2 2 1 Install the cleaning unit hose Note An air hose is required to operate the sensor with the optional cleaning unit An oil free compressed air 6 bar and a solenoid valve or the HOAB compressed air cleaning system is also required refer to section 7 3 on page 27 With the HOAB compressed air cleaning system replace the end of the hose c...

Страница 14: ...14 Installation Figure 7 Install the cleaning unit hose Figure 8 FP 360 sc measurement sensor with cleaning unit and chain mount kit ...

Страница 15: ...5 Installation 3 2 3 Installation with flow cell The FP 360 sc sensor with flow cell is installed for samples free of solids and particulates and limited sample flows Figure 9 Installation with flow cell ...

Страница 16: ...16 Installation ...

Страница 17: ...e corresponding sensor and confirm 4 Select CONFIGURE and confirm 5 Select EDIT NAME and confirm 6 Edit the name and confirm to return to the SENSOR SETUP menu Complete sensor configuration in the same manner with the following menu options selected SET PARAMETER SELECT UNITS AVERAGE LOG SETUP GAIN VALUE 4 3 Sensor data logger There is a data log and event log available for each sensor In the data...

Страница 18: ... 1 to 100 OFFSET 1000 to 1000 FACTOR OFFSET 2 POINT 3 POINT 4 POINT 5 POINT FACTOR Is shown when FACTOR has been selected under CONFIGURE Refer to section 4 5 on page 20 for detailed information OFFSET Is shown when OFFSET has been selected under CONFIGURE Refer to section 4 5 on page 20 for detailed information 2 POINT Is shown when 2 POINT has been selected under CONFIGURE Refer to section 4 5 o...

Страница 19: ... DEFAULT CONFIG AUTO SET DEFAULTS Security query reset to preset configuration for all menu options listed above DIAG TEST PROBE INFO SENSOR NAME Device name EDIT NAME SERIAL NUMBER Device serial number RANGE 0 to 500 or 0 to 5000 MODEL NUMBER Item no Sensor CODE VERSION Sensor software COUNTER OPERATING HOURS Operating hours counter MAINTENANCE Counter counting down days BULB CHANGE Counter count...

Страница 20: ...ti point calibration needs to be carried out Both of these operations must be carried out on site using lab analysis data Basic prerequisites for quantitative measurements are precise knowledge about the oil type in occurrence and constant measurement conditions e g in cool water in a heat exchanger If several oils are present in varying quantities it is generally not possible to carry out a quant...

Страница 21: ... 4 5 3 Check the zero point Medium ultra pure water Target value 1 ppb Clean the window in the event of deviations Use a glass container not plastic large enough to enable the measurement to be taken with an 8 to 10 cm distance between the measurement window and the base e g a 1000 mL glass beaker Place a black non reflective underlay under the container and switch off artificial lights during the...

Страница 22: ...ti point calibration 2 to 5 point calibration Note Multi point calibration means that the pairs must be input in ascending order 1 Open the MAIN MENU 2 Select SENSOR SETUP and confirm 3 Select the corresponding sensor and confirm 4 Select CALIBRATE and confirm 5 Select CONFIGURE and confirm 6 Select the type of calibration e g 2 POINT and confirm 7 Select 1PAIR and confirm 8 Edit the TARGET VALUE ...

Страница 23: ...ction cycles To maintain the performance of the connectors Protect the neoprene components of the connector against high temperatures and intense solar radiation If this cannot be avoided moisten the dried out neoprene components with clean clear water Subsequently loosen the connector The pins must always be lightly lubricated Only apply a very thin film of lubricant Recommended lubricants are Lo...

Страница 24: ...lly remove the deposits on the measurement window Avoid sharp objects on the measurement window surface 3 Rinse the residue from the cleaning agent with fresh water CAUTION Cleaning agents can be hazardous to health Wear protective equipment and avoid direct contact with cleaning fluids NOTICE Other cleaning agents can damage the material Damage caused by cleaning carried out incorrectly is not co...

Страница 25: ...all manufacturer customer service Displayed warnings Definition Resolution DIAG TEST Counter expired Call manufacturer customer service BULB CHANGE Counter expired Call manufacturer customer service LAST CONFIGUR Changed configuration was not accepted Send configuration again TARGET VALUE With multi point calibration values not entered in ascending order Enter calibration values in ascending order...

Страница 26: ...26 Troubleshooting ...

Страница 27: ...itanium housing cable length 10 m without cleaning unit LXV441 99 22101 Range 0 to 500 5000 ppb titanium housing cable length 10 m with cleaning unit and fittings not in connection with flow cell LXV441 99 22201 Range 0 to 500 5000 ppb titanium housing cable length 1 5 m without cleaning unit LXV441 99 22301 7 2 Replacement parts Description Cat no Connector cable 1 5 m LZY623 Connector cable 10 m...

Страница 28: ... parts and accessories Hose for measuring medium 5 m 8 6 mm external internal diameter LZY672 Hose for measuring medium 10 m 8 6 mm external internal diameter LZY673 7 3 Accessories table continued Description Cat no ...

Страница 29: ... Tél 32 0 15 42 35 00 Fax 32 0 15 41 61 20 info hach lange be www hach lange be DR LANGE NEDERLAND B V Laan van Westroijen 2a NL 4003 AZ Tiel Tel 31 0 344 63 11 30 Fax 31 0 344 63 11 50 info hach lange nl www hach lange nl HACH LANGE APS Åkandevej 21 DK 2700 Brønshøj Tel 45 36 77 29 11 Fax 45 36 77 49 11 info hach lange dk www hach lange dk HACH LANGE AB Vinthundsvägen 159A SE 128 62 Sköndal Tel 4...

Страница 30: ...hach lange gr HACH LANGE E P E 27 Avlidos str GR 115 27 Athens Tel 30 210 7777038 Fax 30 210 7777976 info hach lange gr www hach lange gr HACH LANGE D O O Ivana Severa bb 42 000 Varaždin Tel 385 0 42 305 086 Fax 385 0 42 305 087 info hach lange hr www hach lange hr HACH LANGE MAROC SARLAU Villa 14 Rue 2 Casa Plaisance Quartier Racine Extension MA Casablanca 20000 Tél 212 0 522 97 95 75 Fax 212 0 5...

Страница 31: ...ntification of the fault If the customer fails to notify the supplier the product is considered approved despite the defect Further liability for any direct or indirect damages is not accepted If instrument specific maintenance and servicing work defined by the supplier is to be performed within the warranty period by the customer maintenance or by the supplier servicing and these requirements are...

Страница 32: ...32 Warranty and Liability ...

Страница 33: ... sec 40051 Unsigned integer 1 R W Measuring interval from 1 sec to 300 sec 40052 Unsigned integer 1 R W Amplification for the low high range probe 0 AUTO 1 0 01 to 50 500 PPB 2 0 01 to 500 5000 PPB 40055 Unsigned integer 1 R W Set output mode for calibration 40056 Unsigned integer 1 R W Set output mode for service 40057 Float 2 R Version of application file 40059 Float 2 R The entry is for the app...

Страница 34: ...34 Modbus register ...

Страница 35: ...spection interval 6 M Maintenance schedule 23 Measurement accuracy 5 Measuring method 5 Measuring principle 8 Measuring range 5 Modbus register 33 P Precautionary labels 7 Pressure range 5 Product contents 8 R Reproducibility 5 Response time 5 S Safety information 7 Safety sleeve 11 Specifications 5 System configuration 17 W Warnings 25 Wearing parts 25 ...

Страница 36: ...36 Index ...