gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 20
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
5.
Remove the 4 feet of the treadmill.
12.2. Treadmill support assembly
Date:
Name:
Time:
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
6.
Screw the load cells on left and right foot
plates. Sensor 1 on front left, clockwise
numbering, and cables on back side.
Torque =
50Nm
4x DIN7991-M12x40
Note: Ensure that each load cells is
tightened at its nominal torque on the
first mounting side before tightening the
second
mounting
side,
to
avoid
damaging the insert on load cell support
part.
7.
Attach the cable glands with locknuts on the
4 Capteur-Mini3D-Case-Open. Let them
slide freely on the cable.
4x Capteur-Mini3D-Case-Open