11.6 Safety inspection

In order to maintain the safety of the device, h/p/cosmos prescribes performance of an annual safety inspection.
Refer to the date on the inspection sticker on your device for the next inspection date.

h/p/cosmos bases the annual safety inspection on  German laws and regulations.
It is the operator’s responsibility to comply with national laws and regulations.   

The inspection label on the device also certifies the optional equipment and the accessories.
However, inspection intervals for optional equipment and accessories may deviate.

After 12 months or 5000 km there will be a safety inspection reminder (see below).

The error message will be reset by the h/p/cosmos service personnel performing the safety inspection.

11.7 Spare parts and consumables

Spare parts must only be replaced by qualified h/p/cosmos service personnel.
Information about spare parts is available at 

[email protected]

.

A list of consumables is included in the accompanying documents.

11.5 Issues for qualified service personnel

All maintenance work that is not explained in detail, must not be performed by the operator.
Safety inspections, installation and repair work, must also not be performed by the operator.
This kind of work must be performed by h/p/cosmos service personnel according to the “h/p/cosmos service instructions”.
The “h/p/cosmos service instructions” are available at 

[email protected]

.

Содержание 170-190/65 3p Series

Страница 1: ...ign software and intellectual property is owned by h p cosmos sports medical gmbh except where explicitly stated differently Warning All rights reserved Unauthorized copying reproduction hiring lending public performance and broadcasting prohibited quasar models stellar lt stellar quasar lt quasar stellar lt med stellar med quasar lt med quasar med pulsar lt pulsar locomotion 190 65 E med locomoti...

Страница 2: ...s 170 190 65 Models stellar lt med cos30003va17 stellar med cos30003va18 quasar med cos30003va20 quasar lt med cos30003va19 pulsar lt cos30004va01 pulsar cos30004va03 quasar med 3p cos30003va26 pulsar lt 3p cos30004va02 pulsar 3p cos30004va04 Product family treadmill h p cosmos 170 190 65 3p Models stellar lt cos30003va13 quasar lt cos30003va15 stellar cos30003va14 quasar cos30003va16 locomotion 1...

Страница 3: ...c research Many developments and pioneering work from h p cosmos have influenced not only product design and functionality but also their usage and methodologies Your success with our devices is the primary goal of h p cosmos This is why we offer individual devices as well as comprehensive system solutions You will find a wide range of options and accessories in these instructions for use and at w...

Страница 4: ...ded duration 13 3 5 Contraindications 14 3 Intended Use sport 15 3 1 Intended use 15 3 2 Intended operator 15 3 3 Intended location 15 3 4 Intended duration 15 3 5 Contraindications 16 4 Safety med 17 4 1 Safety information Forbidden use 17 4 Safety sport 19 4 1 Safety information Forbidden use 19 4 2 Fall prevention devices 21 4 3 Emergency dismount 23 4 4 Emergency off 24 4 5 Emergency stop 24 4...

Страница 5: ...terfere with automatic program 40 8 8 Pause function 41 8 9 Acceleration levels 42 8 10 User Options 43 9 Accessories compatible devices 45 9 1 Creating Systems 45 9 2 Overview of accessories 45 9 3 Compatible devices 46 10 Disinfection Cleaning 47 11 Maintenance 48 11 1 Intervals and competences 48 11 2 Daily inspection 48 11 3 Lubrication 49 11 4 Adjustment of running belt 50 11 5 Issues for qua...

Страница 6: ... pole disconnection 58 13 Technical data 59 13 1 UserTerminal 59 13 2 Dimensions 59 13 3 Loads 59 13 4 Emissions 59 13 5 Essential performance characteristics 60 13 6 Environmental conditions 60 13 7 Technical and legal requirements 61 13 8 EMC tests 61 13 9 Classification 61 13 10 Certificates 62 13 11 Interfaces RS232 D SUB 9 pole 62 13 12 Voltage Current Performance 62 13 13 Software Programs 6...

Страница 7: ...tting cos15933 cos100548 cos15294 83 Special speed 0 to 10 km h 25 km h 30 km h 40 km h 45 km h 84 Heart rate measurement POLAR non coded cos102818 85 Heart rate measurement POLAR W I N D coded cos100106 86 Heart rate measurement POLAR H10 cos101787 01 87 Arm support adjustable cos12013 88 robowalk expander cos30022 cos30023 89 optofix cos103386 90 Safety arch cos10079 01 91 Handrail long 2 pillar...

Страница 8: ...page 8 of 99 cos105000_170 190_65 rev 1 03 1988 2019 h p cosmos sports medical gmbh ...

Страница 9: ... warning danger warning or caution statements DIN EN ISO 7010 W001 Warning of obstacles stumbling DIN EN ISO 7010 W007 Warning of electrical voltage DIN EN ISO 7010 W012 Warning of hot surface DIN EN ISO 7010 W017 Warning of counter rotating rollers trapping zones DIN EN ISO 7010 W025 Follow instructions for use DIN EN ISO 7010 M002 Potential equalization IEC 60445 Protection ground IEC 60417 5019...

Страница 10: ...y up ISO7000 0623 Keep dry ISO7000 0626 Centre of gravity ISO7000 0627 Temperature limitations SO7000 0632 Do not stack ISO7000 2402 1 2 Symbols used transport 1 3 Structure Illustration Description sport Pages marked with the following symbol are valid for medical devices only sport Pages marked with the following symbol are valid for sports devices only med sport ...

Страница 11: ...l rear panel UserTerminal 5B outs cos11933 01 label h p cosmos address cos10144 01 Umrandung nur Schnittmuster nicht mitdrucken Laufgurteinstellung Adjust running belt Anleitung lesen Read manual COS10512 L R www h p cosmos com COS10512 aufkleber laufgurteinstellung ps M _Artikeldetails Zeichnungen COS10512 aufkleber laufgurteinstellung cdr Mittwoch 22 Januar 2003 22 46 33 Farbprofil Generisches C...

Страница 12: ...unning deck enables the elevation Both parameters are manually adjustable on the UserTerminal Furthermore operation is possible via pre and self defined modes Operation is also possible via external devices PC ECG etc The chapter operation gives a detailed description of all functions The chapter technical data shows technical details The treadmill is driven by powerful motors For that reason it i...

Страница 13: ...ct of a medical application is not necessarily a patient Therefore these instructions for use will use the term subject for patients as well as for athletes under test h p cosmos medical treadmills may be operated with healthy subjects as well For applications with healthy subjects please apply the instructions for use for sports devices available at www hpcosmos com 3 2 Intended operator Medical ...

Страница 14: ...n has to be made by the medical doctor before the treadmill is used Left main coronary stenosis Main artery disease Cardiac valve disease of moderate severity Known electrolyte imbalance Arterial hypertonia RR 200 mm Hg syst 110 mm Hg diast Tachyarrhythmia or bradyarrhythmia Hypertrophic cardiomyopathy and other forms of outflow tract obstruction Higher degree atrioventricular AV blocking Anemia P...

Страница 15: ...stop in the following cases while performing sprints for max endurance tests while training on running surfaces wider than 65 cm for children 14 years for subjects with all kind of disabilities impairments visual hearing balance etc activity limitations and participation restrictions while running with shoes with spikes or studs during reverse belt rotation or use the pull cord safety stop actuato...

Страница 16: ...of moderate severity Known electrolyte imbalance Arterial hypertonia RR 200 mm Hg syst 110 mm Hg diast Tachyarrhythmia or bradyarrhythmia Hypertrophic cardiomyopathy and other forms of outflow tract obstruction Higher degree atrioventricular AV blocking Anemia Physical and or mental disabilities leading to inability to exercise adequately Partially invasive medical devices probes infusions cathete...

Страница 17: ...immediately and consult a medical doctor Exclude overloading or overstressing of the subject The subject has to be checked by a medical doctor before using the device A defibrillator must be present at any time The intended operator has to be in reach of at least one emergency stop off at any time Obey all information given in these instructions for use Do not use the device against the intended u...

Страница 18: ...chnical data Do not use the device without instruction by authorized personnel acc to the instruction protocol Regard safety area behind device of 2 0 m x width of treadmill Operator and subject have to be aware of automatic load changes during profile cardio and test mode Unmeant trapping hazards Take off ties scarfs or other clothes that may be trapped Secure long hair and ribbons during mainten...

Страница 19: ...brillator must be present at any time The intended operator has to be in reach of at least one emergency stop off at any time Obey all information given in these instructions for use Do not use the device against the intended use Do not use the device in case one or more of the listed contraindications prevail Neither subject nor operator must be under the influence of alcohol drugs or anesthetics...

Страница 20: ...me room with the device Operator and subject have to be aware of automatic load changes during profile cardio and test mode Unmeant trapping hazards Take off ties scarfs or other clothes that may be trapped Secure long hair and ribbons during maintenance and training in order to prevent being captured in trapping zones Perform a daily visual inspection see chapter maintenance Obey the maintenance ...

Страница 21: ...ty arch or a body weight support device airwalk It is up to the operator to use any other device that prevents the subject from falling and complies with IEC60601 1 The pull cord safety stop is not a fall prevention Do not use the safety harness on bare skin Treadmill with safety arch Body weight support device airwalk Further information see Annex III accessories Description Ilustration Put on th...

Страница 22: ...ion Ilustration Press lever to release rope Check rope daily Renew the rope and harness every 24 months or sooner if it shows any signs of wear or damage Select the user position on the treadmill using the safety arch In order to re open the buckle press the fastener with thumb and index finger ...

Страница 23: ...bles and falls into fall prevention device Treadmill stops Operator subject hits the emergency off Operator helps subject stand up again Operator helps subject exit the device Subject lost consciousness and is hanging in the fall prevention device Hit the emergency off Call a medical doctor Call one or more persons strong enough to carry the subject Inform the third person that you will open the b...

Страница 24: ...l stop button Do not stand on or enter the running deck when device is in elevation running belt might slip through due to gravity Operation Result Release Restart UserTerminal Power supply to load generating components interrupted Running belt spins out until 0 km h with undefined deceleration Movement of elevation system stops UserTerminal off Interface communication interrupted Switch the devic...

Страница 25: ...l These risks are reduced by safety information within the IFU It cannot be excluded that unintended or forbidden use might cause further not yet regarded risks and that already regarded risks might have been estimated incorrectly It can also not be excluded that the daily use of the medical product might show further risks For medical applications such as ergometry diagnostics and therapy there a...

Страница 26: ... prevention Guide subject onto treadmill Do not enter the device when running belt is rotating Do not step on rear roller Do not stand on or enter the running deck when device is in elevation running belt might slip through due to gravity If possible the subject should hold both handrails for stability when entering the treadmill running belt might slip through due to gravity Holding handrails dur...

Страница 27: ...ord so that the subject has to maintain the correct position see position of subject and operator 5 1 Further preparation locomotion Description Ilustration Adjustment of handrails Pull lever Push button Adjust handrails Release button Push lever Adjustment of therapist seats Release fixation Adjust seat Tighten fixation Adjustment of foot rest Release fixation Adjust foot rest Tighten fixation ...

Страница 28: ... mode or UKK walk test The displays may show service information and error messages as well see trouble shooting adjust units and decimals with OP 12 14 Element Primary function Secondary functions Select mode Decrease increase current speed Navigate through settings adjust parameters Decrease increase current elevation See interfere with automatic program Conform mode Start the operation Confirm ...

Страница 29: ...as standard or lt light devices lt devices have no UserTerminal no display no keyboard Standard device lt device Control via UserTerminal Software see accessories additional keyboard see accessories ECG spirometry etc see compatible devices Control via Software see accessories additional keyboard see accessories ECG spirometry etc see compatible devices ...

Страница 30: ...rfaces are located at the rear of the UserTerminal Unused RS232 interfaces must be covered with dust caps cos102973 for isolation purposes Description Ilustration RS232 COM1 Possible connection of external devices standard RS232 COM2 Possible connection of external devices optional RS232 COM3 Possible connection for service only optional RS232 COM4 Possible connection of external devices optional ...

Страница 31: ...emergency off button at the UserTerminal body height obstacles etc the operator must install an additional emergency stop within reach see accessories 3 Training area acc to ISO 20957 1 subject device 4 Free area acc to ISO 20957 1 training area 0 6 m must be free at all times except operator 5 Safety area acc to DIN EN 957 6 2 0 m behind device must be free at all times except operator 6 Patient ...

Страница 32: ...e device is switched on all emergency offs are released Switch the running machine on with the on off switch light goes on Make sure the space under the treadmill is free from persons body parts or objects especially when switching on treadmill will lower during initialization and when changing the elevation Normal condition When starting all displays show 0 values Select an operation mode For a d...

Страница 33: ...le Manual mode Select mode Confirm and start mode Select speed Select elevation Terminate application Cardio mode Select mode Confirm and start mode Adjust parameters indicated by LEDs Confirm profile Terminate application Profile mode Select mode Confirm and start mode Select profile Confirm profile Terminate application Test mode Select mode Confirm and start mode Select test Confirm test Termin...

Страница 34: ... predefined starting speed default 0 5 km h To predefine starting speed see OP09 in User Options To request body weight for calculation of energy and power see OP16 in User Options Confirm with enter Running belt starts with predefined starting speed default 0 5 km h Pressing and holding the key accelerates with defined acceleration level see acc levels Decelerate to 0 km h to pause the operation ...

Страница 35: ... 0 5 10 15 20 25 0 5 10 15 0 5 10 15 20 elev in v in km h t in min profile 2 5 0 5 10 15 20 25 0 5 10 15 0 5 10 15 20 elev in v in km h t in min profile 3 6 Profiles 1 and 4 profile 1 without elevation step v in km h t in min elev in 1 6 5 04 00 0 2 9 0 00 30 5 3 6 5 03 00 0 4 9 0 00 30 10 5 6 5 03 00 0 6 9 0 00 30 10 7 6 5 03 00 0 8 9 0 00 30 10 9 6 5 03 00 0 10 9 0 00 30 10 11 6 5 03 00 0 12 9 0...

Страница 36: ... is displayed Select profile with or Current profile is displayed in elevation display Max speed is displayed in speed display Duration is displayed in time display Confirm with enter The selected profile starts with the first profile step after countdown The operation stops automatically after the last step To scale the profiles activate OP11 User Options If scaling is performed the max parameter...

Страница 37: ...nals beeps Furthermore the displays show the next load parameters flashing The cardio mode allows training within pre defined heart rate limits In order to stay within the limits the treadmill adjusts speed and elevation automatically first speed then elevation A POLAR heart rate sensor detects the heart rate Moisten the contact areas of the POLAR heart rate sensor Place the transmitter directly u...

Страница 38: ...m The treadmill adjusts speed and elevation according to following matrices Deviation from lower limit Speed km h Elevation Reaction time s 5 heart beats 0 2 0 1 25 6 15 0 4 0 2 25 16 30 0 6 0 4 25 31 50 0 8 0 8 20 50 heart beats 1 0 1 0 20 Deviation from upper limit Speed km h Elevation Reaction time s 5 heart beats 0 3 0 3 12 6 15 0 8 0 8 12 16 30 1 0 1 0 10 31 50 1 5 1 2 8 50 heart beats 2 0 1 ...

Страница 39: ...ation Select test with or Confirm with enter The running belt starts automatically after countdown Following tests require further settings 01 UKK walk test 02 Graded test 03 Conconi test 10 Ramp profile Terminate the operation with stop For possibilities to interfere with an automatic program see interfere with an automatic program ...

Страница 40: ...This sets a new max speed Decelerate to 0 km h to pause the operation see pause function Alter elevation Press up or down Profile test mode Only valid for current step Switch between modes Press enter together with or Switch to automatic mode Further settings are necessary to continue Profile or test mode Switch between steps Press enter together with up or down Cardio mode only Change heart rate ...

Страница 41: ...t stops The speed display shows PAUS Pressing the start key starts the running belt at the preset starting speed default 0 5 km h All values are maintained Pressing the key starts the running belt at 0 1 km h All values are maintained Pressing the stop key one time terminates the application All values remain on the display for 2 minutes Pressing the stop key a second time resets all values ...

Страница 42: ...acceleration level Example Pressing 3 times then holding results in an acceleration with acceleration level 3 Acceleration levels 1 4 are freely accessible Acceleration levels 5 7 are locked by the administrator options For access please contact our service department The high acceleration of the levels 5 7 is dangerous for untrained subjects and must only be applied during sports medicine and ath...

Страница 43: ...rts medical gmbh Basic functions Buttons displays Further information Press and stop together for at least 3 s speed display shows OP 01 Select option with or Confirm with enter Adjust option with or Confirm with enter Leave options by pressing cancel 8 10 User Options ...

Страница 44: ...em 0000 0000 all senders accepted xxxx xxxx only specific sender accepted 9999 9999 next sender will be accepted OP 20 RS 232 interface protocol COM 1 OFF 1 20 1 h p cosmos coscom OP 21 RS 232 interface protocol COM 2 OFF 1 18 1 h p cosmos coscom OP 23 RS 232 interface protocol COM 4 OFF 18 23 20 h p cosmos coscom v3 OP 27 Min acceleration and deceleration level 1 5 OP 28 Max acceleration and dece...

Страница 45: ...cessories are available for this device For illustrations and detailed descriptions see annex or www hpcosmos com Article number Accessory Purpose Information cos10079 01 h p cosmos safety arch Fall prevention and safety stop Mandatory cos10071 v4 1 0 Para control PC software Remote control software Included cos00097010034 Interface connection cable RS 232 5m Connection Included cos10223 Potential...

Страница 46: ...oard left extra wide speed 170 Speed training Optional cos14764 Footboard left extra wide speed 190 Speed training Optional cos14042 Floor protection mat treadmill 170 65 Floor protection stability Optional cos14043 Floor protection mat treadmill 190 65 Floor protection stability Optional cos12598 Base plate 170 65 for h p cosmos airwalk Floor protection stability Optional cos12606 Base plate 190 ...

Страница 47: ...ion Illustration Disinfection Unplug the device Apply some disinfectant to a cloth Wipe all surfaces the subject might have touched Wipe all surfaces that may have come into contact with body fluids Wipe the safety harness h p cosmos recommends Bacillol plus order number cos12179 Clean the device regularly Unplug the device Use a damp cloth not wet Wipe all surfaces Wash safety harness acc to labe...

Страница 48: ...ompetence Daily inspection Daily Operator Lubrication When OIL message occurs Operator Adjustment of running belt If due Operator Safety inspection 12 months h p cosmos service personnel only Change of safety arch rope 24 months h p cosmos service personnel only Tightening of the running belt If due h p cosmos service personnel only Adjustment of levelling sockets If due h p cosmos service personn...

Страница 49: ...Do the same if there are dry grinding noises during operation The OIL messages occurs every 1000 km by default The oil interval may be varied depending on environmental conditions and use Unplug the treadmill Move one hand tissue for 1 m through the center between running belt and running deck A slight film of oil must remain on the hand If the running belt is too dry it must be lubricated For fur...

Страница 50: ... to the right Turn counter clockwise to adjust belt to the left After adjustment observe the running belt at 10 km h for at least 2 min Belt must maintain the position Remove Allen key from screw 11 4 Adjustment of running belt Unmeant trapping hazards Take off ties scarfs or other clothes that may be trapped Secure long hair and ribbons during maintenance and training in order to prevent being ca...

Страница 51: ... After 12 months or 5000 km there will be a safety inspection reminder see below The error message will be reset by the h p cosmos service personnel performing the safety inspection 11 7 Spare parts and consumables Spare parts must only be replaced by qualified h p cosmos service personnel Information about spare parts is available at service hpcosmos com A list of consumables is included in the a...

Страница 52: ...top In case there is no improvement contact service hpcosmos com Elevation does not work E21 Switch off Wait 10 min in order to cool down Switch on again In case E21 still appears unplug device exclude reconnection mark clearly and inform h p cosmos service personnel via telephone and writing Device does not start but displays speed value Switch off and on again Check error messages see below Oil ...

Страница 53: ...ia RS232 blocked COM port Restart peripheral software Restart peripheral device Acceleration via external device too slow Check max acceleration and deceleration level via RS232 User Options OP 29 Any other problem unplug device exclude reconnection mark clearly and inform h p cosmos service personnel via telephone and writing Error code Acoustic code x short beep o long beep Error message Action ...

Страница 54: ...nal Running surface dimensions L 170 x W 65 cm L 190 x W 65 cm Track access height 23 cm Handrail dimensions Ø 6 cm Ø 4 cm Data may be influenced by accessories 13 3 Loads Description Data stellar cos30003 quasar 3p cos30003 pulsar 3p cos30004 locomotion 3p cos30024 Max subject weight 300 kg Device weight 330 kg 335 kg 355 kg 345 kg 385 kg 560 kg 600 kg Substitutional load to floor EN 1991 3 2 kN ...

Страница 55: ...y cell phones containers with liquid etc can fall into the device or onto the running surface or underneath the running belt Operation Operation Data Temperature 10 50 C Humidity 30 70 without condensation Pressure 700 1060 hPa Altitude max 3000 m without pressurization altitudes 1000m can cause minor loss of performance Oxygen saturation 25 Central lubrication system manual Exclude presence of ex...

Страница 56: ...ield immunity test EN 61000 4 8 Voltage dips short interruptions and voltage variations immunity tests EN 61000 4 11 Testing of voltage changes voltage fluctuations and flicker in public low voltage supply systems EN 61000 3 3 Variation of mains frequency DIN EN 60601 1 13 9 Classification Description Data MDD 93 42 EEC Notified body MDD 93 42 EEC Risk class IIb active therapeutic device and activ...

Страница 57: ...lar n a polar 13 12 Voltage Current Performance Description Data stellar cos30003 quasar cos30003 pulsar cos30004 locomotion cos30024 quasar 3p cos30003 pulsar 3p cos30004 locomotion 3p cos30024 Input voltage 230 V f 50 60 Hz 400 V f 50 60 Hz Current input long time 7 0 A 6 8 A Current input momentary 15 2 A 14 3 A Energy consumption long time 1600 VA 2700 VA Energy consumption momentary 3500 VA 5...

Страница 58: ...osmos para motion non medical optional gaitway non medical optional Noraxon non medical optional Microgate Optogait optional zebris Rehawalk optional Number of stored programmes 6 training profiles 10 test profiles 8 user definable profiles For other coscom compatible software see website www coscom org 13 14 All pole disconnection The following options are available for all pole disconnection Unp...

Страница 59: ...65 cm Track access height 23 cm Handrail dimensions Ø 6 cm Data may be influenced by accessories Lower profile versions or custom made devices on request 13 3 Loads Description Data stellar cos30003 quasar cos30003 Max subject weight 300 kg Device weight 310 kg 315 kg Substitutional load to floor EN 1991 3 1 kN m 3 1 kN m Load on each support wheels feet 1 9 kN 1 9 kN Data may be influenced by acc...

Страница 60: ... into the device or onto the running surface or underneath the running belt Operation Operation Data Temperature 10 50 C Humidity 30 70 without condensation Pressure 700 1060 hPa Altitude max 3000 m without pressurization altitudes 1000m can cause minor loss of performance Oxygen saturation 25 Central lubrication system manual Exclude presence of explosive infammable acid and corrosive gases Exclu...

Страница 61: ...turbances induced by radio frequency fields EN 61000 4 6 Power frequency magnetic field immunity test EN 61000 4 8 Voltage dips short interruptions and voltage variations immunity tests EN 61000 4 11 Testing of voltage changes voltage fluctuations and flicker in public low voltage supply systems EN 61000 3 3 Variation of mains frequency DIN EN 60601 1 13 9 Classification Description Data MD 2006 4...

Страница 62: ...te 115200 bps n a polar 13 12 Voltage Current Performance Description Data stellar cos30003 quasar cos30003 Input voltage 230 V f 50 60 Hz Current input long time 7 0 A Current input momentary 15 2 A Energy consumption long time 1600 VA Energy consumption momentary 3500 VA Energy efficiency N A Device protection switch circuit breaker 16 A Drive motor capacity 3300 W Elevation motor capacity No el...

Страница 63: ...tway non medical optional Noraxon non medical optional Microgate Optogait optional zebris Rehawalk optional Number of stored programmes 6 training profiles 10 test profiles 8 user definable profiles For other coscom compatible software see website www coscom org 13 14 All pole disconnection The following options are available for all pole disconnection Unplug device from power socket Unplug cable ...

Страница 64: ...com Limited liabilities apply If h p cosmos or h p cosmos organizational bodies senior management or agents can be held accountable for the payment of damages pertaining to slight negligence breach of material contractual obligations the damages shall be limited to damages that could typically have been foreseen Liability pertaining to slight negligence excludes liability as a result of loss of pr...

Страница 65: ... bolt If in compliance with local national standards use type A RCDs residual current operated protective device For type B RCDs measuring also direct current an additional leakage current limiter has to be installed in the treadmill The intended location must provide the requirements for electrical installation acc to technical data The intended location must maintain the local requirements for e...

Страница 66: ...re and connectors are explained in detail UserTerminal 10 The general use is explained incl manual profile cardio and test mode Operation 11 The competences and intervals for maintenance are explained Maintenance 12 The adjustment of the running belt is explained in detail Maintenance 13 The lubrication of the device and owner s observation duties are explained in detail Maintenance 14 The accompa...

Страница 67: ...y information intended or forbidden use as well as unauthorized maintenance or lack of maintenance and regular safety checks may lead to injury or even death and can damage the device Furthermore this will result in loss of liability and warranty Please fill out the instruction protocol and send it back to h p cosmos by fax 49 18 05 16 76 69 email sales hpcosmos com or mail h p cosmos device model...

Страница 68: ...stance covered in km indicates km OP03 Indication of total hours of operation stand by time including runtime of motor running belt h and show Operation hours reports h OP04 Indication of total hours of runtime of motor running belt h and report Operation hours shows h OP05 Indication of firmware version and date reports OP05 reports typE indicates device type e g 1 4 indicates MCU 5 reports versi...

Страница 69: ...ation of power and energy consumption OP16 Request for body weight before manual or automatic start 0 OFF Request for body weight before starting a program is not required Calculation of energy consumption and power is based on the body weight entered in option no 15 1 ON Input of body weight before starting a program is required Calculation of energy consumption and power is based on the entered ...

Страница 70: ...s For advanced h p cosmos coscom v3 connections with baud rate 115200 please select OP20 20 Please note that the connected devices software must be approved for h p cosmos coscom v3 with baud rate 115200 for example h p cosmos para control 4 1 OP21 RS232 interface protocol COM 2 See descriptions above OP23 RS232 interface protocol COM 4 indicates flashing Number of the RS232 interface protocol and...

Страница 71: ...he h p cosmos coscom protocol the selected level in option 29 or in the corresponding command of the h p cosmos coscom protocol OP40 Locking and unlocking of the treadmill OFF After switching on the treadmill is completely locked not accessible To unlock the treadmill press the buttons and START simultaneously While locked the display shows no ACCESS ON treadmill is unlocked accessible standard OP...

Страница 72: ...after pressing START again and will not be automatically deleted by pressing STOP Display values can only be deleted by pressing the STOP key twice time distance energy OP46 Report mode display Elevation in profile mode and test mode 0 Display alternates default 1 Display permanently in or degree depending on OP14 2 Display permanently in Step After total switch off the default value 0 will be val...

Страница 73: ...rforming the UKK walk test the subject must warm up and determine the max walking speed During the test the subject must walk as fast as possible for 2 km heart rate approx 80 of max The subject must not run The treadmill measures the heart rate via polar heart rate measurement every 500 m Affter performing the test the UKK Fitness Index is displayed The UKK walk test is suitable for subjects betw...

Страница 74: ...y the medical doctor Skip remaining break time Press start once restart after countdown Press start twice restart immediately Prolong break Press within break pause is indicated Press start to continue test continues with remaining break time Description Illustration e g for performance diagnostics based on heart rate measurement Refer to test mode for safety information and adjustment Endurance t...

Страница 75: ...on 1 3 00 2 7 10 2 4 0 12 3 5 4 14 4 6 7 16 5 8 0 18 6 8 8 20 7 9 6 22 Description Illustration e g for ECG stress test Regard chapter test mode for safety information and adjustment Step Duration min Speed km h Elevation 1 3 00 3 0 0 0 2 3 5 3 7 0 4 10 5 5 14 0 6 17 5 Description Illustration e g for ECG stress test Refer to test mode for safety information and adjustment Step Duration min Speed ...

Страница 76: ... test mode for safety information and adjustment Step Duration min Speed km h Elevation 1 3 00 2 7 10 0 2 4 8 10 0 3 6 4 10 0 4 8 0 10 0 5 8 0 15 0 6 9 6 15 0 18 7 Cooper protocol Description Illustration e g for ECG stress test Refer to test mode for safety information and adjustment Start at 5 3 km h and 0 elevation After 1 minute elevation increases to 2 After another minute the elevation is in...

Страница 77: ... steps onto the running belt the doctor presses the START key again By pressing the START key the second time the displays will be reset to zero Step Duration min sec Speed km h Elevation Total time min sec 0 until START is pressed 3 2 0 until START is pressed Test phase Subject steps onto the running belt 1 02 00 3 2 0 2 00 2 02 00 3 2 2 4 00 3 02 00 3 2 4 6 00 4 02 00 3 2 6 8 00 5 02 00 3 2 8 10...

Страница 78: ...r for at least 5 s Select speed for step 1 with or Confirm with enter Select distance for step 1 with or Confirm with enter Select distance 0 in order to program this step by time Select time for step 1 with or Confirm with enter Select time 0 00 in order to program this step by distance Select elevation for step 1 with or Confirm with enter 18 12 Self defined tests Tests 21 40 are freely definabl...

Страница 79: ...splays Further information Select acceleration for step 1 with or Confirm with enter The acceleration level is displayed in the index display Proceed the same way for further steps Make sure all steps after the last one have a speed of 0 km h Press start for at least 5 s to save the program and exit programming ...

Страница 80: ... the PC and keys can be simulated for remote control of the running machine or the ladder ergometer Illustration Application See separate instructions for use Additional safety information See separate instructions for use Technical data Min processor Pentium IV OS Windows XP Vista 7 RAM 1 GB 2 GB recommended Free HD 200 MB Resolution 1280 x 1024 Microsoft NET Framework 3 5 Service Pack 1 Microsof...

Страница 81: ...elease Restart Push button Running belt stops with predefined deceleration Movement of elevation system stops UserTerminal displays pull stop Mains connection and interface communication not interrupted Release button Restart application Additional safety information N A Technical data N A Additional accessories N A Installation By h p cosmos service personnel only Further information https www hp...

Страница 82: ...Application The additional keyboard allows full control over all functions of the MCU like START STOP elevation UP and DOWN speed and Additional safety information See safety information for the treadmill Technical data Connection plug 7 pin male screwed L x W x H 20 x 8 x 2 cm Cable coiled cable extendable up to 2 meters Additional accessories cos100816 Extension cabel for h p cosmos additional k...

Страница 83: ...ement 10m cos15294 L15m Emergency stop ext without attachement 15m Illustration with magnet holder without attachement Application Operation Result Release Restart Push button Running belt stops with predefined deceleration Movement of elevation system stops UserTerminal displays pull stop Mains connection and interface communication not interrupted Release button Restart application Additional sa...

Страница 84: ...hort description Extends max speed to 10 km h 25 km h 30 km h 40 km h 45 km h Illustration N A Application N A Additional safety information N A Technical data cos15258 cos12995 cos11995 cos10158 cos10159 max 10 km h max 25 km h max 30 km h max 40 km h max 45 km h Additional accessories N A Installation By h p cosmos service personnel only Further information N A ...

Страница 85: ...an POLAR T31 or T34 is used see print If the heart rate is not displayed or displayed incorrectly There might be interferences with Screens computers printers mobile phones and any radio engineering systems Electric devices electric motors transformers High voltage transmission lines also from trains Strong fluorescent tubes nearby Central heating radiators Other electric devices In order to preve...

Страница 86: ...ing immediately and consult a medical doctor Technical data Transmission radius approx 10m Further data see accompanying POLAR documents or www polar com Troubleshooting In case the heart rate is not displayed Chest belt might be applied incorrectly see application above Other chest belt than POLAR W I N D is used see print Device and chest belt are not connected see User Option OP 19 Additional a...

Страница 87: ...y stop exercising immediately and consult a medical doctor Technical data Further data see accompanying POLAR documents or www polar com Troubleshooting In case the heart rate is not displayed Chest belt might be applied incorrectly see application above Other chest belt than POLAR W I N D is used see print Device and chest belt are not connected see User Option OP 19 Additional accessories cos100...

Страница 88: ...orts outside of training area when running Do not use on bare skin Do not leave the arm support in a position that projects into running area Before loading make sure the adjustment elements are correctly locked Do not use the arm supports with reverse belt rotation Technical data Adjustability Height and width via 3 joints Measurements 48 x 42 5 x 26 cm each packed Weight 10 7 kg each Max subject...

Страница 89: ...rox 110 cm depending on treadmill Weight approx 15 kg depending on treadmill Max pulling force 50 N per rope Technical data robowalk back cos30023 Height approx 80 cm depending on treadmill Weight approx 25 kg depending on treadmill Max pulling force 50 N per rope Additional accessories cos101051 XS leg cuff XS shank for circumference 14 27 cm cos101050 S leg cuff S thigh for circumference 25 39 c...

Страница 90: ...the treadmill The fastening profiles can be fixed without tools only by magnetic force The option is independent of the frame height Do not step on optofixTM Additional safety information Do not step on the fixed bars or fastening profiles In case there is no optogait optojump the optofix has to be removed not usable with mounted footboard Technical data Length 12 cm Width 8 6 cm Height 14 9 cm Ad...

Страница 91: ... prevention devices Additional safety information See safety information for the treadmill Technical data Max subject weight 300 kg Max subject height 200 cm Min triggering force 30 kg Min room height 250 cm treadmill at 0 elevation 275 cm treadmill at 25 elevation Additional accessories cos14903 03 XXS chest belt system XXS chest circumference 45 65 cm cos14903 03 XS chest belt system XS chest ci...

Страница 92: ...old handrails for balance control Holding handrails during use affects exercise results Additional safety information N A Technical data Lenght Diameter Weight 170 stellar quasar 224 5 cm 6 cm 18 kg each 190 pulsar 243 5 cm 6 cm 19 kg each Additional accessories N A Installation By h p cosmos service personnel only Further information https www hpcosmos com en products supplies options handrails b...

Страница 93: ...load Be careful with squeeze and shear points Pay attention to the right position of the handrails Before loading make sure the adjustment elements are correctly locked Technical data adjustability handrail tube length tube diameter weight material 150 50 height 55 120 cm width 38 104 cm 150 cm 4 cm 69 kg steel tube powder coated 170 190 65 height 75 120 cm width 62 128 cm 170 190 cm 4 cm 71 kg st...

Страница 94: ... Illustration Application Subject may hold handrails for balance control Holding handrails during use affects exercise results Additional safety information N A Technical data Lenght Diameter Height Weight 132 cm 6 cm 98 5 cm 12 kg each Additional accessories N A Installation By h p cosmos service personnel only Further information https www hpcosmos com en products supplies options handrail left ...

Страница 95: ...e without supervisory staff Do not switch during motion Max speed for reverse belt rotation is reduced by default Technical data speed range 0 5 0 km h 3 1 mph in reverse mode no limitation in regular mode speed limit for reverse mode can be undone in case a fall prevention is used Modification of standard belt is replaced by reinforced belt basic device with profiled surface and 5 mm thickness be...

Страница 96: ...nal balance control Illustration Application Subject may hold crossbar front rail for balance control Holding handrails during use affects exercise results Additional safety information It is not allowed to run with the back to the crossbar Technical data N A Installation By h p cosmos service personnel only Further information https www hpcosmos com en products individual products handrail crossb...

Страница 97: ...a Lenght Width Weight 170 stellar quasar 167 cm 40 cm 22 kg each 190 pulsar 187 cm 40 cm 25 kg each Additional accessories Separable handrails cos14233 02 are necessary Installation By h p cosmos service personnel only Further information https www hpcosmos com en products supplies options footboard right extra wide speed 17065 https www hpcosmos com en products supplies options footboard left ext...

Страница 98: ...ake off ties scarfs or other clothes that may be trapped Secure long hair and ribbons during maintenance and training in order to prevent being captured in trapping zones The ramp must not touch the running belt Make sure the ramp cannot slip Always enter from the back not from the side Do not install the ramp when running belt is in motion Technical data Length Width Height Weight 150 50 124 cm 8...

Страница 99: ...rman mobile networks fax 49 18 05 16 76 69 email service hpcosmos com Skype hpcosmos com search select name Sales phone 49 18 05 16 76 67 0 14 min from German landlines max 0 42 from German mobile networks fax 49 18 05 16 76 69 email sales hpcosmos com Skype hpcosmos com search select name h p cosmos sports medical gmbh Am Sportplatz 8 DE 83365 Nussdorf Traunstein Germany phone 49 18 05 16 76 67 0...

Отзывы: