10.4 Relay 4 (RLY4).
After the instrument has been switched on , or the 1/3 MHz-key has
been pressed, the
µ
p checks the head sizes of all heads connected.
RLY4 closes the HEAD_SIZE line, and reads the size of all heads
connected to the apparatus.
After reading, RLY4 opens again. This is done in order to obtain a
galvanic separation of
the outputs in case of combination therapy.
10.5 Microprocessor (
µ
p) pin description.
•
USD_UP_KEY Input signal. Switches to logic low level when
the ultrasonic
density up key is pressed.
•
USD_DOWN_KEY
Input signal. Switches to logic low level when
the
ultrasonic density down key is pressed.
•
SELECT_LEFT_KEY Input signal. Switches to logic low level when
the left
output mode key is pressed.
•
SELECT_RIGHT_KEY Input signal. Switches to logic low level when
the right
output mode key is pressed.
•
N.U.
Not used.
•
FREQ_ 1_3_MHZ_KEY
Input signal. Switches to logic low level when
the 1/3 MHz.
key is pressed.
•
TIME_DOWN_KEY
Input signal. Switches to logic low level when
the time
down key is pressed.
•
TIME_UP_KEY
Input signal. Switches to logic low level when
the time
µ
p
key is pressed.
•
WATCH_DOG
Output signal. In case of
µ
p hang up, this
signal is at
logic high level.
•
KEY_SELECT
Output signal. Logic low enables hardware
information to
be read in the
µ
p. Logic high level disables
hardware
information and connects the keyboard to the
µ
p.
•
VUS_EN
US voltage enable. Low/high active. Disables
output stage
in case of
µ
p hang up.
•
SCK
Serial clock. Clock signal for I•C- bus.
•
SDA
Serial data. Data signal for I•C- bus.
•
SCK_ADDA
Clock signal for I•C -bus, connected only to U15
(ADDA)
•
SDA_ADDA
Data signal for I•C-bus, connected only to
U15(ADDA)
•
HEAD_SIZE
Input signal. This line is read during
initialising, to let the
µ
p
know the head sizes of the heads connected.
•
SERVICE_REQUEST
High active. When a jumper is placed on CON1 ,
this line
switches to logic high level and the service
routine is
activated.
•
ON/OFF
Output signal: When this signal is at logic high
level, the
1/3 MHz signals are cut off, which results in 0
W output.
This signal will switch on/off at 100 Hz in
case of
combination therapy.
•
1/3 MHZ
Output signal. When this signal is at logic high
level, the
3 MHz base frequency is selected.
Page 14