OPERATION
167
OCP Test Automation
Background
The OCP test function creates an automatic test to
test the OCP of power supply products.
Parameters
Active Channel
Applies the setting to the
load channel.
Range
High(CC Mode High) or
Low(CC Mode Low)
Start Current(Start C) Starting current value for
the test.
End Current(End C) The current value that will
end the test. The value
must be higher than the
OCP value of the DUT you
are testing. This parameter
is used as a fail-safe for if
the over current protection
of the DUT fails.
Step Current(Step_C) Sets the step resolution of
the current.