LCR-6000 Series User Manual
27
4.1.2 Impedance Range [RANGE]
Impedance range mode
To achieve the best measurement result, it is
recommended to set the impedance range mode to
“Auto range”. When using “Hold range”, errors in
measurement results may occur if the measured value
exceeds the full scale of the selected range.
Table 4-4
Impedance range mode
Mode
Function
overview
Advantage
Disadvantage
Auto
Range
LCR-6000
Series sets the
optimum
impedance
range for the
impedance of
the DUT
You don’t
need to select
range
The
measurement
time is longer
due to the
ranging time
Hold
Range
Measurement
is performed with a fixed impedance range
No ranging
time is
required
You need to
select a proper
range
depending on
the value of
the DUT.
Nominal
Range
LCR-6000
Series sets the
optimum range
depending on
the nominal
value
You don’t
need to select
range. No
ranging time is
required
Valid only in
the sorting
mode
Available impedance range
The hold range has nine impedance ranges: 10, 30, 100,
300, 1k, 3k, 10k, 30k and 100k.
The impedance range is selected according to the DUT’s
impedance even if the measurement parameter is
capacitance or inductance.