![GW Instek GSM-20H10 Скачать руководство пользователя страница 213](http://html1.mh-extra.com/html/gw-instek/gsm-20h10/gsm-20h10_user-manual_2280064213.webp)
GSM-20H10 User Manual
211
Example
:CALCulate2:LIMit2:LOWer DEFault
Command :CALCulate2:LIMitx:LOWer?
Function
:LOWer? Query specified lower limit.
:LOWer? DEFault Query *RST default lower limit.
:LOWer? MINimum Query lowest allowable lower
limit.
:LOWer? MAXimum Query largest allowable lower
limit.
Example
:CALCulate2:LIMit2:LOWer?
Command :CALCulate2:LIMitx:UPPer[:DATA] <n>
Function
This Command is used to set the upper limits for
LIMIT 2, LIMIT 3, and LIMIT 5 through LIMIT 12
tests. The actual limit depends on which measurement
function is currently selected. For example, a limit
value of 1μ is 1μA for the amps function and 1μV for
the volts function. A limit value is not range sensitive.
A limit of 2 for volts is 2V on all measurement ranges.
LIMitx x=2, 3, 5-12
<n> = -9.999999e20 to 9.999999e20 Specify limit value
DEFault Set specified upper limit to 1
MINimum Set specified limit to -9.999999e20
MAXimum Set specified limit to +9.999999e20
Example
:CALCulate2:LIMit2:UPPer DEFault
Command :CALCulate2:LIMitx:UPPer?