GSM-20H10 User Manual
209
Function
This Command automatically acquires the null offset
value. If no reading is available, then the next available
reading will become the null offset value.
Example
:CALCulate2:NULL:ACQuire
Command :CALCulate2:NULL:STATe <b>
Function
This Command is used to enable or disable null offset.
When enabled,the CALC2 reading will include the
null offset value. When disabled, CALC2 will not
include the null offset.
<b> = 1 or ON Enable null offset
0 or OFF Disable null offset
Example
:CALCulate2:NULL:STATe 1
Command :CALCulate2:NULL:STATe?
Function
Query state of null offset.
Example
:CALCulate2:NULL:STATe?
Command :CALCulate2:DATA?
Function
This Command is used to acquire all the readings used
for the CALC2 limit tests. NOTE that if null offset is
enabled, then the CALC2 readings will include the null
offset value. At least one of the limit tests have to be
enabled to acquire limit test readings.
Example
:CALCulate2:DATA?
Command :CALCulate2:DATA:LATest?
Function
This Command operates exactly like CALC2:DATA?,
except it returns only the latest null offset or limit
result.
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...