COMMAND SYNTAX
208
<name> = CALCulate[1] Use result of CALC1
VOLTage Use measured voltage
reading
CURRent Use measured current
reading
RESistance Use measured resistance
reading
Example
:CALCulate2:FEED VOLTage
Command :CALCulate2:FEED?
Function
Query input path for limit tests.
Example
:CALCulate2:FEED?
Command :CALCulate2:NULL:OFFSet <n>
Function
This Command lets you establish a null offset (REL)
for the selected feed. When Null Offset is enabled the
result is the algebraic difference between the feed
reading and the offset value:
CALC2 reading = feed reading - null offset.
<n> = -9.999999e20 to 9.999999e20 Specify null
offset value
Example
:CALCulate2:NULL:OFFSet -9.999999e20
Command :CALCulate2:NULL:OFFSet?
Function
Query null offset value.
Example
:CALCulate2:NULL:OFFSet?
Command :CALCulate2:NULL:ACQuire