ASR-3000 Series User Manual
88
Setting the Output Off Phase
Background
The off phase setting sets the ending phase of the
voltage output.
Steps
1.
Press
Shift + 4
to make the OFF Phs
parameter selectable.
+
2.
Also, it is available to use the
scroll wheel followed by the Enter
key to make the OFF Phs
parameter selectable as well.
3.
Set the
OFF Phs
setting with the scroll wheel or
with the F3 (MAX) and F4 (MIN) soft-keys to set
the Off Phase to the maximum and minimum
values, respectively.
AC+DC-INT, AC-INT, AC+DC-ADD,
AC-ADD, AC+DC-Sync, AC-Sync, AC-VCA
OFF Phs
Range
0.0° ~ 359.9°
Soft-keys
FIXED/FREE, MAX, MIN
4.
Press
Enter
to confirm the Off Phase setting.
Example of OFF
Phase Setting
OFF Phase setting
F1
F3
F4
Soft-keys setting