OPERATION
79
Setting the Output Frequency & Signal
Background
The FREQ and SIN settings set the frequency of the
output. Before setting the frequency, set the
frequency limit.
Steps
1.
Press the
F
key to access the FREQ
or SIG parameter depending on
varied modes.
Also, it is available to use the scroll
wheel followed by the Enter key to
make the FREQ or SIG parameter
selectable as well.
2.
Set the frequency or signal with the scroll wheel
or with the F1 ~ F4 soft-keys.
AC+DC-INT, AC+DC-ADD
FREQ
Range
1.00 ~ 999.9 Hz
Soft-keys
DEF1, DEF2, MAX, MIN
AC-INT, AC-ADD, AC-VCA
FREQ
Range
40.00 ~ 999.9 Hz
Soft-keys
DEF1, DEF2, MAX, MIN
AC+DC-Sync, AC-Sync
SIG
Option
LINE, EXT
3.
Press
Enter
key to confirm the frequency or
signal setting.
Defined Settings
The DEF1 and DEF2 settings are user defined
settings. By default they are set to 50.00 Hz and
60.00 Hz, respectively. The MAX and MIN soft-keys
set the frequency to the maximum and minimum,
respectively.
Содержание ASR-2000 Series
Страница 1: ...Programmable AC DC Power Source ASR 2000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 11: ...GETTING STARTED 11 ...
Страница 130: ...ASR 2000 Series User Manual 130 Example Power ON setting ...
Страница 216: ...ASR 2000 Series User Manual 216 ASR 2000 Dimensions ASR 2100 2050 Scale mm ...
Страница 217: ...APPENDIX 217 ASR 2100R 2050R Scale mm ...
Страница 218: ...ASR 2000 Series User Manual 218 ASR 2100R 2050R with GET Series Box Scale mm ...
Страница 219: ...APPENDIX 219 GET 003 Series Box GET 004 Series Box ...