PEL-3000 Series Parallel Assembly Guide
12
IV.
Based on the connection method in the figure below, connect
current sink cable to the unit followed by setting current sink
value for unit. Measure the currents from both the current sink
cable of Master positive pole and the current sink cable of Booster
positive pole, via current probe, to validate the accuracy of
currents. The measured current value of Master is supposed to be
1/3 of the set value, whereas the measured current value of
Booster should be 2/3 of the set value.