Calibrating the 40T
The
bbnoisecal
script automatically performs appropriate averaging to
reduce the effects of aliasing and cultural noise.
5.3.1
Sensor response codes
Sensor
Sensor type code
Units
(V/A)
40T-1 or 6T-1, 1 s – 50 Hz response
CMG-40_1HZ_50HZ
V
40T-1 or 6T-1, 1 s – 100 Hz response
CMG-40_1S_100HZ
V
40T-1 or 6T-1, 2 s – 100 Hz response
CMG-40_2S_100HZ
V
40T-1 or 6T-1, 10 s – 100 Hz response
CMG-40_10S_100HZ
V
40, 20 s – 50 Hz response
CMG-40_20S_50HZ
V
40, 30 s – 50 Hz response
CMG-40_30S_50HZ
V
40, 60 s – 50 Hz response
CMG-40_60S_50HZ
V
5.4
Calibration with a hand-held control unit
If you prefer, you can inject your own calibration signals into the system through a
hand-held control unit. The unit includes a switch which activates the calibration
32
Issue F - June 2019