9
©
Guldmann 1
1/2021 # 5617
31_101
©
Guldmann 1
1/2021 # 5617
31_101
2 .00 Labels and Marking
CE marking
Medical Device Class I in accordance
with EU MDR Regulation
Read the manual before use
Caution
Product label
xxxxxx
xxxxxx
V. Guldmann A/S
Graham Bells Vej 21-23A
8200 Aarhus N, Denmark
yyyy-mm-dd
max xxx kg / xxx lbs
1: 5
max
xxx cm
xxx inch
3 .00 Maintenance and Storage
3 .01 Cleaning
Clean the ramps with a high pressure cleaner, a damp cloth,
sponge or soft brush applied with a normal cleaning agent.
The ramps may also be rinsed under running water from garden
hoses or the like.
Chemicals and autoclaving may not be used to clean the ramp.
3 .02 Storage
The ramps should be stored in a room where the humidity does
not exceed 70%. The ramps must not be stored in bathrooms.
3 .03 How to Prevent/Avoid Corrosion
The ramp is made from corrosion-resistant materials, and does
not require further corrosion prevention.
3 .04 What Maintenance Is Required By the Owner?
The hinges on the Stepless folding ramps should be lubricated
with TRI-FLOW or a similar lubricating agent every six months, or
as required.
The sliding rails of the Stepless telescopic ramps should be dried
off with a dry cloth when needed.
If the ramps are slippery (covered with ice), use salt on the ramp
- NEVER gravel.
The telescope system must not be taken apart by the user
If the telescope system of the telescopic ramp is taken apart, it
should be taken apart and put together again by qualified person
-
nel.
Tightening moments:
Hinge = 10 Nm. Other = 5 Nm.
Safety: All screws and nuts should be fastened with Locktite no. 270.
3 .05 Disposal
Local and national regulations on environmentally correct recy-
cling must be observed.
4 .00 Transport
4 .01 Locking of telescopic ramps during transport
4 .02 Locking of wide folding pro ramps during transport
5 .00 Service and lifetime
5 .01 Lifetime
Given the maintenance prescribed above, Stepless ramps will
have a useful life of 15 years.
Spare part lists and drawings can be ordered from the
manufacturer or the distributor.
5 .02 Safety/service inspections
Parts which are worn or defective should be replaced by new
spare parts from V. Guldmann A/S.
1 . Visual check of the ramp
• Check the ramp for wear and tear.
• Check the ramp for any faults.
• Check that there is no other form of damage to the ramp.
2 . Testing of the ramp, as for normal use
• Check all the ramp’s functions, loaded and unloaded.
3 . Control of the mechanical state of the ramp
• Clean dirt and other impurities from the ramp.
• Inspect and assess vital parts of the ramp.
• Replace any parts of the ramp which are defective or worn.
Содержание Stepless 30000-050
Страница 35: ...35 Guldmann 11 2021 561731_101 ...