1. SAFE TY RULES

  

This meter is designed for indoor use at temperatures between 0°C to 40°C and 
altitudes up to 2,000m.

  To ensure that the meter is used safely, follow all safety and operating instructions in 

this operation manual. If the meter is not used as described in this operation manual, 
the safety features of this meter might be impaired.

   Do not use the meter if the meter or test leads look damaged ,or if you suspect that 

the meter is not operating properly.

   When using the instrument, keep your fi ngers behind the fi nger guards on the plastic 

casing and probes.

   Disconnect the live test lead before disconnecting the common test lead.

   Make sure power is off before cutting, desoldering, or breaking the circuit wires. Small 

amounts of current can be dangerous.

  Do not apply more than 600 VDC or 600V AC rms between a terminal and ground.

   To avoid electrical shock, use CAUTION when working above 60V DC or 25V AC 

rms. Such voltages pose a shock hazard.

   Never make measurements with the battery cover off.

  To avoid electrical shock or damage to the meter, do not exceed the input limits.

2. INTERNATIONAL SYMBOLS

 

  

Important information  

 Dangerous Voltages 

 

see manual  

  Continuity

 

  AC  

 Ground

 

  DC 

   Double Insulation

3. TECHNICAL SPECIFICATIONS

3.1 General Specifi cations

Display:  

3 3/4 digit LCD with decimal point and  

 

33 segments bargraph display.

Polarity: 

Automatic, (-) negative polarity indication

Zero adjustment:  

Automatic

Sample rate: 

0.5 Sec.

Over range indication:   Symbol 

 is displayed

Power:  

2 x 1.5 V battery, type AA, UM3 or equiv.

Power saving:  

Automatic power off

Battery life: 

Approx. 50 hours. (w/ alkaline batteries)

Dimension:  

6.34 x 3 x 2” (HxWxD).

Weight:  

Approx. 11.4 Oz. (including batteries)..

Accesories: 

User’s Manual,Test Leads,Protective,  

 

Holster,  Soft  Pouch and  2 x 1.5 Volt      

 alkaline 

batteries 

3.2 Electrical Specifi cations

Accuracies are ±(% of r number of least signifi cant 
digits) at 23°C ±5°C, less than 75% RH.

4. OPERATION

4.1 Instrument Description

4.2 LCD Display Description

Function

Range

Accuracy

Input 

Impedance

Remarks

Overload 

Protection

DC Voltage

320 mV

0.8%+2

2 M

-

600 Vp-p

3.2V, 32V, 

320V, 600V 

1.2%+2

AC Voltage

3.2V, 32V, 

320V, 600V

1.2%+5

2 M

50~400Hz

600 Vp-p

DC Current

320

A, 

3200

A, 

32mA,320mA

1.5%+2

< 163 mV Voltage Drop

10A max. for <10 Sec

Fuse 

protection 

10 A - 250V, 

10A max. 

10 sec.

10A

2.5%+2

AC Current

320

A, 

3200

A, 

32mA,320mA

1.5%+5

< 163 mV Voltage Drop

10A max. for <10 Sec

Fuse 

protection 

10 A - 250V,  

10 A max 

10 sec.

10A

2.5%+5

Function

Range

Accuracy

Remarks

Overload 

Protection

Resistance

320

, 3.2K

32k



,320k



3.2M



1.2%+5

Max. Testing Voltage:

< 1.5 V

250 Vp-p

32M

3.5%+5

250 Vp-p

Continuity

Buzzer sounds when <18

Test Voltage: 1.2 V Max.

250 Vrms

Diode Test

Displays forward diode Volt-

age drop (Vf)

Test Current < 0.6mA

250 Vrms

4.3 Measurement Procedures

 

 

CAUTION

:  Maximum Input Voltage is 600Vrms,do not exceed 

this rating to avoid personal injuries or damage to 
the instrument. The range FUNCTION/RANGE 
switch should be set to the range you want to test 
before the operation.

  

CAUTION

:  Always ensure that the correct terminals are used for 

the type of measurement to be made. Avoid making 
connections to “live” circuits whenever possible.
When making current measurements ensure that 
the circuit is not “live” before opening it in order to 
connect the test leads.

OFF

V

V

Ω

μ

A

mA

10A

AUTO   

SELECTION

AC  DC

max 10A

fused

10A

COM

CAT II

MAX 600V

400mA fused

AUTO OFF

V

Ω

mA

max

600V

HOLD

RANGE

MAN
mV
Mk

+

-

H

0

.........

1

.........

2

.........

3

..

GTC 

CT8017

1

2

3

7

6

4

5

MAN
mV
Mk

+

-

H

0

.........

1

.........

2

.........

3

..

Beeper enable

Digital Readout

Manual range
enable

Measurement
units

Low battery

Analog bar graph

DC

Data Hold

AC

Diode test

enable

1) Case
2) Display
3) Hold button
4) Function switch
5) Input and common 

connectors

6) Alterate function button

7) Range button

CT

8017

 