USER GUIDE
Gas Sensing Solutions Ltd.
Revision 1.2, 29 March 2021
Page | 23 For regular updates, sign up at https://gassensing.co.uk
Copyright © 2021 Gas Sensing Solutions Ltd.
PRESSURE COMPENSATION
GSS sensors are calibrated at a nominal 1013mbar. The CO
2
reading from the sensor will vary from
the nominal output if the mean barometric pressure is different from the calibration setting.
It is possible to configure the sensor to correct for this effect by writing the actual or reference mean
barometric pressure in mbar to the sensor. This can be done as part of the initial set up process or
updated at any time during use. Write the decimal value in mbar into the dialogue box and click
‘Pressure Comp’.
In all cases the sensor polling mode needs to be switched off during the change. The software will
warn the user if this is not selected. After the change has been made polling mode can be re-
enabled.