10. Screw the additional pH and redox or temperature electrodes
into the designated
locations (pos. A1/A2 or D). Only place
the potentiostatic measuring electrode (pos. E), do not screw
it in.
11. Connect the measuring cables and electrodes with the same
marking colour.
12. Contact the counter-electrode (pos. M) with the measuring
amplifier (pos. P/Q).
13. Connect the sample water inlet (pos. J) and outlet (pos. K).
14. Open the sample water inlet (pos. J), and adjust the flow with
the adjusting spindle (pos. L).
15. Connect the cleaning motor (pos. H) with the power supply.
16. Screw in the potentiostatic measuring electrode (pos. E)
manually up to the stop.
17. Before calibration, keep a running-in period of at least two
hours.
For further maintenance, cleaning and repair, follow the
instructions given in the installation and operating
instructions of the components.
14. Accessories
14.1 Included accessories
• Buffer solution (1 container)
• 2 spare tubes for peristaltic pump.
14.2 Spare parts
14.2.1 Spare parts for AQC-D5 preassembled system
14.2.2 Spare parts for measuring cell AQC-D5
Caution
Do not touch the electrode surface (pos. E / 1)!
Caution
Please observe also section
Caution
Strictly observe the maintenance instructions in the
installation and operating instructions of the
components!
Description
Product
number
Buffer solution
91835799
(549-300)
Spare parts set for peristaltic pumps
(containing 1 tube)
91836486
(553-1100)
Description
Product
number
Spare parts set for the potentiostatic measuring
cell consisting of 15 balls (glass), magnetic
stirring stone, slide disk and gasket
91835822
(553-1010)
Spare electrode for the potentiostatic measuring
cell
91835242
(314-251)
Spare parts set for the sample water filter
consisting of spare filter element with gasket and
adaptor
91835368
(48.876-1)