background image

H31.0.11.6C-11

User’s Manual GMH3110

page 6

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

  

7  Error And System Messages

Display

Meaning

What to do?

Low  battery power, device will only continue
operation for a short period of time

Replace battery

Battery empty

Replace battery

Mains operation without battery: wrong voltage

Check  power supply, replace it when
necessary

No sensor connected

Switch off device and connect sensor

   

or 

Err.9

Connected sensor or device defective

If second sensor available, check if device
is ok. Return defective device/sensor to
manufacturer for repair

Err.9

Measured value far out of allowable range

Check: pressure not within sensor range?

Battery empty

Replace battery

Mains operation without battery: wrong voltage
or polarity

Check  power supply, replace it when
necessary

System error

Disconnect battery and power supplies,
wait shortly, then reconnect

No display or

confused

characters,

device does
not react on

keypress

Device defective

Return to manufacturer for repair

Measured value above allowable range

Check: pressure not within sensor range?
-> measuring value to high

Err.1

Sensor defective

Return to manufacturer for repair

Measured value below allowable range

Check: pressure not within sensor range?
-> measuring value to low

Err.2

Sensor defective

Return to manufacturer for repair

Err.4

Value is too low to be displayed, tara is set

Check: display below -2000 (tara?)?

Err.7

System error

Return to manufacturer for repair

Содержание GMH3110

Страница 1: ...H31 0 11 6C 11 User s Manual Handheld Digital Pressure Meter GMH3110 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...densation may cause in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to...

Страница 3: ...he left side of the device 1 4 Display 1 5 Basic Operation When switching on the device and a zero point adjustment was carried out it shows shortly nuLL Corr On Off Switch Tara Calling of tara functi...

Страница 4: ...P oFF Auto Power Off Time The device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off time The power off time can be set to...

Страница 5: ...ce and a suitable electrically isolated interface adapter GRS3100 GRS3105 or USB3100 the device can be connected to a computer for data transfer With the GRS3105 up to 5 devices of the GMH3xxx series...

Страница 6: ...sured value far out of allowable range Check pressure not within sensor range Battery empty Replace battery Mains operation without battery wrong voltage or polarity Check power supply replace it when...

Страница 7: ...o values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7 mm high for measuring values and for min max memories hold function etc as well as additiona...

Страница 8: ...BAE 0 0 400 0 bar abs 0 1 bar 600 bar abs C Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1...

Отзывы: