Operating Manual GLF100

page 3 of 6

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

2   General Information About Conductivity Measuring

2.1  The Conductivity Measuring Cell

During the measurement, the conductivity measuring cell must be dipped at least in so far, that at least 30mm
beginning from the top of the measuring cell, is located in the medium.
The maximum immersion depth for continuous operation should not exceed 110mm.

The measuring cell can either be stored dry or in water. After dry storage wetting time will be prolonged slightly. If
changing over from one liquid to another with conductivities varying widely make sure to properly rinse and shake dry
measuring cell.

Attention: Measuring cell must never come into contact with water-repellent materials such as oil or silicone.

If conductivity measured is much higher or lower than expected this may be due to the electrode being soiled with non-
conducting or conducting foreign materials. Measuring cell has to be cleaned with a watery soap solution.
When measuring media with low conductivities the electrode has to be stirred sufficiently.

2.2 Measuring 

hints

Conductivity measuring is comparably easy to perform, the precision of the instrument is very constant if it is used as
intended. Depending on the necessary accuracy the instruments can be used up to several years without recalibration
of the cell constant.
If the accuracy  should be controlled or improved, this is done by means of suitable reference solutions and the
adjusting of the cell factor.
Attention! Wrong handling of reference solution can make them useless very fast.

Measuring procedure:

Especially when measuring low conductivity: Before immersion to the measuring solution, rinse the electrode with
deionised water, dry with a paper towel and shake remaining water out.

The measuring is speeded up considerably, if the electrode is immersed and pulled out the solution several times.
Especially when measuring low conductivity the electrode needs sufficient flow during the measuring, e.g via stirring
the solution.

When temperature compensation is activated, the electrode needs enough time to adjust to the actual temperature of
the measuring solution to work accurate.

2.3 Temperature 

compensation

The conductivity of aqueous solution is temperature dependent. The dependency itself is strongly dependent on the
kind of solution. For the most applications e.g. in fish farming etc., the non linear temperature compensation of natural
waters is precise (“nLF” according to EN 27888). The most common reference temperature is 25°C.

2.4  Min-/max- value memory

The lowest and highest measured value since the turning-on of the device will be stored.
Watch MIN value (Lo): 

press shortly key 

mode 

once

display changes between ‘Lo’ and MIN value

Watch MAX value (Hi): 

press shortly 

mode 

once again 

display changes between ‘Hi’ and MAX value

restore current value: 

press shortly 

mode 

once again 

current value of the chosen unit is displayed

delete MIN/MAX value: 

press 

mode 

for 2 sec. 

MIN and MAX values will be deleted, the

 

display shows shortly ‘CLr’ (clear)

After switching-off and turning-on again all min-/max- values will be deleted.

2.5  Hold – function

When the 

hold 

– key is pressed, the current measured value will be ‘frozen’ (display symbol: HLD) until the key is

pressed again. Then the device will measure in its normal way again.

Содержание GLF-100

Страница 1: ...UCTIVITY MEASURING 3 2 1 THE CONDUCTIVITY MEASURING CELL 3 2 2 MEASURING HINTS 3 2 3 TEMPERATURE COMPENSATION 3 2 4 MIN MAX VALUE MEMORY 3 2 5 HOLD FUNCTION 3 3 CONFIGURATION OF THE INSTRUMENT 4 4 ADJUSTING OF THE INSTRUMENT 5 5 SYSTEM MESSAGES 5 6 INSPECTION OF THE ACCURACY ADJUSTMENT SERVICES 5 7 SPECIFICATION 6 precise measurements of conductivity TDS total dissolved solids and Salinity Applica...

Страница 2: ...ergency stop device or in any other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 1 2 Operating And Maintenance If the symbol LO BAT is displayed at the left side of display the battery is weak measuring can be continued for a short period If bAt is di...

Страница 3: ...hem useless very fast Measuring procedure Especially when measuring low conductivity Before immersion to the measuring solution rinse the electrode with deionised water dry with a paper towel and shake remaining water out The measuring is speeded up considerably if the electrode is immersed and pulled out the solution several times Especially when measuring low conductivity the electrode needs suf...

Страница 4: ...ng of TDS 0 2000 mg l 50 0 SAL Measuring of Salinity 0 0 50 0 g kg Unit of temperature display C temperature is displayed in C factory setting F temperature is displayed in F temperature compensation only for µS cm and mS cm p r t note below oFF No temperature compensation of conductivity measuring nLF non linear temperature compensation for natural water acc to EN27888 DIN 38404 to measure ground...

Страница 5: ...emperature 2 0 2 0 C 3 6 3 6 F Offset setting of temperature measurement see below Scale of temperature 5 00 5 00 Scale adjustment of temperature measurement Display measured value offset 1 scale adjustment 100 Pressing again stores the settings the instruments restarts segment test Please note If there is no key pressed within the menu mode within 60 seconds the adjusting will be cancelled the en...

Страница 6: ... 100 C Relative humidity instrument 0 95 r F non condensing Storage temperature 5 50 C Housing Dimension 110 x 67 x 30 mm L x B x D impact resistant ABS membrane keyboard transparent panel Front side IP65 Weight approx 155 g Power Supply 9V Battery type IEC 6F22 included Power Consumption 1 5mA Display ca 11 mm high 4 digit LCD display with additional segments Pushbuttons 3 membrane keys for on of...