background image

H55.0.31.6C-05 

Operating Manual GDH 200-14  

page 5 of 8 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

8  Min-/max- value memory 

Display MIN-value (Lo): 

Press button mode shortly 

Display changes between „Lo‟ and min-value  

Display MAX-value (Hi): 

Press mode shortly again 

Display changes between „Hi‟ and max-value 

Display actual value again: Press mode shortly again 

Actual value is displayed 

Delete MIN/MAX: 

Press button mode for 2 s 

MIN&MAX are deleted. „CLr‟ (Clear) is displayed  

  

 

shortly 

The Min-/max- display is automatically set to actual value display after 10 s. 

The Min-/max- values are deleted if the device is switched on after it had been turned off. 

9  Zero-function (relative measuring) 

With help of the Zero-function relative measurements can be realized: press “zero“-button for 2s.  

 NULL

“ is displayed shortly and the displayed value is set to 0. Press again “zero“ for 2s to display 

the absolute value again.  

Note: Setting and resetting the Zero function will delete your min-/max- value memory. 

Field of application: e.g. tyre pressure measurement, pneumatic measuring up to 10 bar rel. 

10 General information on absolute pressure measurement 

 The device measures the absolute pressure. This must not be confused with the pressure at sea 

level display by weather stations. There the pressure loss due to height above sea level is 

deducted. The device is also able to perform this sea-level correction on the displayed air pressure. 

Therefore 

 (Sea Level correction) has to be activated (= “on”) and the current height above 

sea level (

ALTI

 = Altitude in [m])  has to be entered at the configuration. 

 

 

Содержание GDH 200-14

Страница 1: ...H55 0 31 6C 05 from Version 1 1 GDH 200 14 Operating Manual Vacuum Baro Manometer...

Страница 2: ...3 SAFETY INSTRUCTIONS 3 4 OPERATION NOTES 3 5 OPERATION 4 5 1 DISPLAY ELEMENTS 4 5 2 PUSHBUTTONS 4 6 STARTING 4 7 PRESSURE CONNECTION 4 8 MIN MAX VALUE MEMORY 5 9 ZERO FUNCTION RELATIVE MEASURING 5 10...

Страница 3: ...and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it 1 Trouble free operation and reliability of th...

Страница 4: ...tivates deactivates the Zero function 6 Starting Switch the device on with button After the segment test the device displays some information to its current configuration OFFS if offset correction is...

Страница 5: ...off 9 Zero function relative measuring With help of the Zero function relative measurements can be realized press zero button for 2s NULL is displayed shortly and the displayed value is set to 0 Press...

Страница 6: ...20 min 1 120 Auto Power Off turn off delay in minutes If no key is pressed for the time adjusted in this parameter the device is automatically switched off to safe battery power adjustable range 1 to...

Страница 7: ...f there is no key pressed within the menu mode within 2 minutes the adjusting will be cancelled the entered settings are lost 13 Notes to the calibration services Calibration certificates DKD certific...

Страница 8: ...ue is set to zero Display approx 13 mm high 4 digit LCD display Pushbuttons 3 membrane keys for on off min max value request offset adjustment Working conditions 25 to 50 C 0 to 80 RH non condensing S...