Technical data
Measuring range (specified accuracy)
0 .. 2000 ppm
0.000 .. 0.200 %
Measuring range (not specified)
0.000 .. 1.000 %
Accuracy
± 70 ppm ± 3 % measurement
Measuring cycle
2 seconds
Display
3-line segment LCD, additional symbols, illuminated (adjusta-
ble white, permanent illumination)
Additional functions
Min/Max/Hold
TWA calculation / STEL
2-stage alarm (optical and acoustic)
Adjustment
1-point, 2-point and basic sensor adjustment
Housing
Break-proof ABS housing
Protection rating
IP30
Dimensions L*W*H
[mm] and weight
108 * 54 * 28 mm without measuring cell or kink protection
180 g, incl. battery and measuring cell
Operating conditions
0 to 50 °C; 0 to 85 % r.h. (non-condensing)
Storage temperature
-20 to 70 °C
Current supply
2*AA-NiMH batteries (included in delivery)
Current requirement/
Rechargeable battery
life
approx. 50 mA, approx. 60 mA with lighting
Life approx. 24 hours with NiMH batteries (without backlight)
charging time of about 8 hours
Battery indicator
4-stage charge status indicator,
Charge indicator for low charge level: "BAT LO"
Plug connector
Micro USB socket (not a data connection)
Auto-power-OFF function
The device switches off automatically if this is activated
Directives and standards
The devices conform to the following Directives of the Council
for the harmonisation of legal regulations of the Member
States:
2014/30/EU EMC Directive
2011/65/EU RoHS
Applied harmonised standards:
EN 61326-1:2013 Emission limits: Class B
Immunity according to Table 2
Additional errors: < 1 % FS
EN 50581:2012
The device is intended for mobile use and/or stationary opera-
tion in the scope of the specified operating conditions without
further limitations.