8 | Error and system messages 

G 1910-02 

Error and system messages 

 

Display

 

Meaning

 

Possible

 

causes

 

Remedy

 

----

 

No signal from the 
sensor 

Measurement far 
outside of the meas-
uring range 

Sensor not ready 

CO

2

 concentration 

too high 

Defective sensor ad-
justment 

Sensor defect 

Wait the start-up time of the 
sensor 

Place the product in clean 
outdoor air 

Perform sensor adjustment 

Send in for repair 

----

 

The display value 
could not be deter-
mined 

Sensor not ready 

Sensor defect 

Wait the start-up time of the 
sensor 

Send in for repair 

 

No display, 
unclear char-
acters or no 
response 
when but-
tons are 
pressed 

Rechargeable bat-
tery depleted 

System error 

Product is defective 

Rechargeable bat-
tery depleted 

Error in the product 

Product is defective 

Charge battery 

Replace rechargeable bat-
tery 

Send in for repair 

BAT LO

 

Rechargeable bat-
tery depleted 

Rechargeable bat-
tery discharged 

Rechargeable bat-
tery defect 

Charge battery 

Replace rechargeable bat-
tery 

(AL ERR.1

 

Zero point adjust-
ment defective 

 

Incorrectly meas-
ured CO

2

 concentra-

tion for adjustment 

Expose sensor to a test gas 
with 0 ppm CO

2

 

(AL ERR.2

 

(AL ERR.3

 

Defective gradient 
compensation 

 

Incorrectly meas-
ured CO

2

 concentra-

tion for adjustment 

Incorrect CO

2

 con-

centration 

Expose sensor to a test gas 
with known CO

2

 concentra-

tion 

Enter correct value 

(AL ERR.5

 

Time for stability 
recognition ex-
ceeded 

Stability recognition 
lasts longer than 10 
minutes 

Provide a consistent flow 
with a constant CO

2

 concen-

tration 

ERR.1

 

Measuring range ex-
ceeded 

Measurement too 
high 

Defective sensor ad-
justment 

Sensor defect 

Stay within allowable meas-
urement range 

Perform sensor adjustment 

Send in for repair 

ERR.2

 

Measuring range is 
undercut 

Defective sensor ad-
justment 

Sensor defect 

Perform sensor adjustment 

Send in for repair 

ERR.7

 

Sensor error 

Defective sensor ad-
justment 

Sensor defect 

Perform sensor adjustment 

Send in for repair 

 

 

Содержание G 1910-02

Страница 1: ...Operating manual G 1910 02 Compact CO2 monitor with alarm EN...

Страница 2: ...ng elements 10 4 4 Connections 11 5 Operation 12 5 1 Commissioning 12 5 1 1 Explanation 12 5 2 Configuration 12 5 2 1 Explanation 12 5 2 2 Opening the configuration menu 12 5 2 3 Configuring parameter...

Страница 3: ...Table of contents B H90 0 01 DB2 2 1 3 30 12 Service 29 12 1 Manufacturer 29 12 2 Repairs processing 29 12 3 Sales offices 29...

Страница 4: ...ty and warranty of the manufacturer for damages and consequential dam ages are voided with misuse disregarding this document disregarding safety notices assignment of inadequately qualified technical...

Страница 5: ...tion or a configuration of the product Formula Some instructions include a formula for a general understanding of a configuration programming or a setting of the product Outcome of an action Result co...

Страница 6: ...fety of the product can only be guaranteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury...

Страница 7: ...er to life and health 3 0 Normal exhalation concentration ele vated breathing and pulse rate 2 0 20 000 1 0 10 000 Possible shortness of breath 0 5 5000 TWA Maximum for working condi tions 0 1 0 2 100...

Страница 8: ...rding of the mean value weighted over 8 hours TWA or 15 minutes STEL Monitoring of air quality 2 5 Qualified personnel For commissioning operation and maintenance the relevant personnel must have adeq...

Страница 9: ...ed on switched off and configured and the measurements and parameters can be adjusted and held with the operating elements The product is equipped with an integrated opti cal carbon dioxide sensor In...

Страница 10: ...in display Measurement of the current CO2 value Auxiliary display Display of the mean value Bar graph Visualisation of the CO2 value 4 3 Operating elements On Off button Press briefly Switch on the pr...

Страница 11: ...01 DB2 2 1 11 30 Function key Press briefly Freeze measurement Hold Return to measurement display Call up next parameter Long press 2s Start menu configuration ONF appears in the dis play 4 4 Connecti...

Страница 12: ...about the configuration of the product appears in the display TEST Self test An automatic self test is carried out during system start up The bar indicator for the threshold value is displayed and th...

Страница 13: ...onfiguration the last save value is reproduced on the next start up of the product 5 2 3 Configuring parameters of the configuration menu The following representation shows the available parameters an...

Страница 14: ...alarm is trig gered when the value is exceeded When the pre alarm is triggered it can be muted for 5 minutes To do this press any key The display shows LR AL 1 AL 2 Depending on the setting of the par...

Страница 15: ...ct is switched off 1 Press and hold the Down button 2 Press the On Off button to switch on the product 3 Release the On Off button after 1 second and then the Down button in order to call up the Expan...

Страница 16: ...ee Call up of the expanded settings menu 15 1 Select the desired parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up key and Down key 3 The a...

Страница 17: ...ise use the expected value of the test gas according to the analysis certificate or display value of the reference device The changed value is saved and the Expanded settings menu is closed STOR appea...

Страница 18: ...ter an optical band pass filter is passed an IR sensor measures the level of IR light wherein the CO2 values in the optical path are displayed 6 1 2 Design 1 Gas inlet 2 Gas outlet 3 IR lamp 4 Optical...

Страница 19: ...eries 7 2 Battery 7 2 1 Charge status display If the battery status display blinks the batteries are depleted and must be recharged However the device will still operate for a certain length of time I...

Страница 20: ...ies and dispose of them at a suitable collec tion point NOTE Read the following handling instructions before replacing rechargeable batteries and follow them step by step If disregarded the product co...

Страница 21: ...e air can take place Ensure that the setting of the gradient compensation setpoint SL was entered correctly in the Expanded settings menu If reliable values can no longer be displayed we recommend per...

Страница 22: ...bar indicator blinks 4 You can change the value of the gradient compensation with the up key and down key Otherwise confirm the preadjusted value with the function key The 1 point adjustment is finish...

Страница 23: ...dioxide concentration at the sensor is sta ble before the adjustment is started 3 If a stable correct value is reached the display flashes briefly an acoustic signal sounds and the bar indicator blink...

Страница 24: ...depleted Rechargeable bat tery discharged Rechargeable bat tery defect Charge battery Replace rechargeable bat tery AL ERR 1 Zero point adjust ment defective Incorrectly meas ured CO2 concentra tion...

Страница 25: ...40 C Bring product to room tem perature and restart charg ing process SYS ERR System error Error in the product Sensor defect Switch product on off Replace rechargeable bat teries Send in for repair...

Страница 26: ...ice must not be disposed of with household waste Return it to us freight pre paid We will then arrange for the proper and environmentally friendly disposal Private end users in Germany have the possib...

Страница 27: ...Current supply 2 AA NiMH batteries included in delivery Current requirement Rechargeable battery life approx 50 mA approx 60 mA with lighting Life approx 24 hours with NiMH batteries without backlight...

Страница 28: ...gas 5000 ppm CO2 476699 GZ 19 Adjustment set for G 1910 MiniFlo valve for 12 l gas cylinders quick connect adapter 4 mm closing screw gas bezel 479767 GZ 20 Gas connection for G 1910 Quick connect ada...

Страница 29: ...in our service centre Monday to Thursday from 8 00 to 16 00 Friday from 8 00 to 13 00 GHM Messtechnik GmbH GHM GROUP Greisinger Hans Sachs Str 26 Service Centre 93128 Regenstauf GERMANY Tel 49 940293...

Страница 30: ...e 26000 26999 35000 36999 40000 69999 Email vertrieb west ghm messtechnik de Tel 49 2191 9672 0 Fax 49 2191 9672 40 Post code 70000 97999 Email vertrieb sued ghm messtechnik de Tel 49 9402 9383 52 Fax...

Отзывы: