2
System Specifications
Gas Connection
Span NO
X
gas connection
For 6/4 mm tubing
– max. 1 bar
Air supply filter regulator
1/8” BSP connection – max. 8 bar
Air supply quality
Instrument air quality according to ISO 8573-1,
class 3
Analyzer
See details in the G41 NOx/O
2
Analyzer manual
Analyzing Board
Analyzing board dimension
600 × 500 × 140 mm (H × W × D)
Weight
Approx. 10.0 kg without packaging
Approval & Certificate
Lloyd´s Register Type Approval Certificate
Diffusion Probe
Sensor technology
Heated zirconia type sensor
Measurement range
NO
X
: 0…1500 ppm
O
2
: 0.
0…21 %
Sample temperature
0
…500
C
Probe insert length
Approx. 208
…338 mm
for duct diameters 235
…2800 mm
Mounting type
Welding socket size OD: 70.0 mm L: 190 mm; or
t
hread size: 1½” BSP
Connection for
calibration gas and back-flushing
6/4 mm tubing
Calibration air flow
Approx.
0.5…1 l/m
Diffusion probe dimension
– short
285 x 180 x 475 mm (H x W x D)
Diffusion probe dimension
– long
285 x 180 x 600 mm (H x W x D)
Weight including house
Approx. 6.0 kg without packaging
Содержание G4130 NOX/O2
Страница 1: ...PART NUMBER 02156 G4130 NOX O2 ANALYZING SYSTEM WITH DIFFUSION PROBE MANUAL...
Страница 2: ......
Страница 3: ...Version 1 1 Revision Jan 2019 Part nr 02156 2...
Страница 17: ...System Installation Version 1 1 Revision Jan 2019 Part nr 02156 16 Figure 3 7 Analyzing board dimensions...
Страница 18: ...G4130 NOx O2 Analyzing System 17 Figure 3 8 Piping and connection diagram...
Страница 30: ...29...
Страница 31: ......