117
KMV-3901/KMV-3911
Installation & Service Manual
Troubleshooting Common Issues
Troubleshooting consists of using fault-isolation techniques to narrow down the probable
cause to a specific multiviewer component. Schedule these maintenance operations during
off hours when the system is not in use if possible as troubleshooting can cause outage.
Use the following troubleshooting table to diagnose common issues and take corrective
actions to restore operation.
Symptom
Probable cause
Test
Corrective action
Source signal
input is
unavailable
Signal feed is unavailable Apply a signal from signal
generator test set or
another source to the
multiviewer’s input
Verify what the input status LEDs
show. See
Poor connection
The input status LED is
amber
Trace the cable’s path and ensure
connections at both ends are good
Closed
captioning/XDS /
Dolby E / SCTE /
Loudness does not
work for certain
tiles / video feeds
Insufficient licenses have
been installed.
View license status in
XAdmin for each card.
See
page 101. Valid licenses
have a green dot.
Purchase and install a license for
each card as required. Contact
Technical Support for more
information and see
See
Kaleido Software
upgrade /
downgrade has
failed
A Kaleido Software
installation requirement
was not observed
Review the Kaleido
Software Release Notes
for information about
compatibility,
requirements, limitations,
known issues, and bug
fixes for a given software
version. Confirm that the
multiviewer’s memory
has sufficient capacity for
the upgrade.
Install a software version that is
compatible with your hardware or
upgrade your hardware as required.
See also
The multiviewer’s
time is incorrect
and cannot be set
in XAdmin
Time parameters were
not set in the frame
controller
-
Time is set in the multiviewer’s
frame configuration.
Frame
configuration information is found
in the
Densité 3 frame’s
documentation. See
Layout changes
aren't propagated
properly. Signal
probing and
metadata
extraction doesn't
work properly
All network interfaces
have not been configured
or all necessary network
connections have not
been made to a switch
In XAdmin,
Physical
Connections
, verify the
frame’s connection status
Follow the help instructions in the
right pane to correct any network
configuration issues.
Ensure the frame is properly
connected to a switch. See
Содержание KMV-3901
Страница 20: ...20 Installation Mechanical Installation of KMV 3901 3911 Cards in the Densité Frame ...
Страница 26: ...26 Multiviewer Cabling Commissioning the Multiviewer ...
Страница 40: ...40 Frame Control Panel Operation and IP Network Setup Network Considerations for a Multiviewer Cluster ...
Страница 42: ...42 System Configuration XEdit Installation XEdit Installation ...
Страница 102: ...102 System Configuration Enabling Options ...
Страница 137: ...137 KMV 3901 KMV 3911 Installation Service Manual ...