Functions of Software

 2

9

. Is it possible to connect multiple units?

It is possible. A maximum of 10 units can be connected.

Also, by setting the recording group to the same group, it is possible to control the recording start/stop 

all at once.

By setting multiple loggers to the same recording group, a single file of measurement results can be 

obtained in addition to individual files.

By setting the group data join together to "On" in the recording settings, the data in the group will be 

joined together and output as a CSV file.

10

.

How to store the settings of GL100_GL240_GL840-APS in the main unit.

Display "Connection" from [Connect with main unit], and when you cancel [Connect] of the connected 

GL240, [Memory of setting conditions] is displayed. Select [Yes] to store the setting conditions in the 

main unit. (Some setting conditions are not stored.) 

11

.

How long is the recording time?

It is depends on the amount of free space on your PC's hard disk.

When the SD memory card in the main unit is full, recording in the main unit will stop, but data transfer to 

the PC will continue.

In the case of Off, data recording is performed only on the PC side. The data can be recorded regardless of 

the memory of the main unit.

12

.

Is it possible to display the AB cursor data values of the data played back by GL100_GL240_GL840-APS

at

 

the same time?

By setting the A cursor and B cursor in the "Digital Display" shown on the right side of the "Waveform 

Display" screen, it is possible to display at the same time.

"Digital Display" screen can be displayed simultaneously Minimum value, Maximum value, Average 

value, Peak value etc. 

13

.

Is it possible to change the span and position?

The span and position can be changed by using [Y-axis scale expansion/reduction] and [Move up/

down position] in the lower part of GL100_GL240_GL840-APS. 

To select the channel you want to move or change, click on the channel number in the digital value 

display on the right side.

When CH (ALL) is selected, the setting is reflected on all channels.

Содержание midi LOGGER GL240

Страница 1: ...the device with an external battery using a DC drive cable How many V can be input 10 What is the maximum voltage that can be applied to the logic pulse input section 11 What are the specifications o...

Страница 2: ...is the reason why the temperature is displayed much lower when measuring the temperature with a thermocouple 18 What happens to the recorded data if the power of the device is suddenly turned off 19...

Страница 3: ...at is the update timing of the waveform display on the main unit 15 The main unit shows recording in progress but no data has been recorded 16 What is the ZONE under the MONITOR on the discplay 17 Is...

Страница 4: ...it possible to send mail from GL100_240_840 APS to gmail yahoo mail etc 8 What happens if there is a communication failure between the logger and the PC during recording 9 Is it possible to connect m...

Страница 5: ...voltage Please check the table below for the resolution of each range 3 What is the measurement input impedance in ohms The input impedance is 1M The allowable signal source resistance is less than 30...

Страница 6: ...zone to your region UTC 09 00 for Japan Set the SMTP Authentication Method to SMTP AUTH Set SMTP AUTH to PLAIN Set Same as POP3 to No Set SMTP User Name to gmail com Set SMTP Password to your Gmail pa...

Страница 7: ...inputs The range can be set for each channel individually 3 What kind of measurement is the logic input used for It is for recording the sequence of a relay circuit When a signal is input it is recor...

Страница 8: ...2GB and there is Relay capture function that switches files every 2GB Formatting must be done in FAT32 format 9 When operating the device with an external battery using a DC drive cable How many V can...

Страница 9: ...rminal is pulled up to 5V with 10k An input output cable for GL B 513 is required for alarm output 12 Is it possible to use Y terminal The terminal shape is available since it is an M3 screw terminal...

Страница 10: ...following table 17 What is the standard and transmission distance of the wireless LAN option B 568 EEE802 11b g n The communication distance is about 40m in an obstacle free environment It depends on...

Страница 11: ...there is no On Off setting for this filter and the software automatically removes noise from the AC power supply according to the AC power frequency setting in the GL240 If the voltage or temperature...

Страница 12: ...e data that can be recorded on the SD memory card 2GB Measurement conditions There is a limit to the number of channels that can be set 10ms 1ch 50ms 5ch others 10ch analog input only One file can be...

Страница 13: ...s so do not input a voltage that exceeds the specifications even for a moment Input terminal between and terminals part a of below Maximum input voltage 60Vp p 20mV to 1V range 110Vp p 2V to 100V rang...

Страница 14: ...cy of each range The measurement accuracy of this instrument is based on the following conditions 1 Operating environment 23 C 5 C 2 Aging time of main unit 30 minutes or more after power on Filter se...

Страница 15: ...cation range the product may not start charging In this case turn on the screen saver function or turn off the power before charging Operating environment AC driving 0 to 45 C battery driving 0 to 40...

Страница 16: ...40 operates as an access point master unit in wireless LAN mode When operating as an access point master unit it is possible to connect one GL100 WL and also connect to more smart devices However when...

Страница 17: ...m 13 How to display the results of calculations in real time Real time calculation on the main unit can be performed for inter channel calculation and statistical calculation maximum minimum peak effe...

Страница 18: ...e reason why the temperature is displayed much lower when measuring the temperature with a thermocouple When the measurement is stopped press the MENU key and check if the room temperature compensatio...

Страница 19: ...isplay on and off 21 Why does the Memory Full message appear even though there is free space on the SD memory card and can not start recording TThere is a limit to the number of files that can be stor...

Страница 20: ...pplication development library on the included CD 28 Is it possible to back up data to another SD memory card while recording data to the SD memory card It is possible Set the backup interval in Backu...

Страница 21: ...red on Ch 1 and Ch 2 and display it on Ch 3 Also can I use the value as an alarm setting Display is possible using the inter channel calculation function However it is not possible to set alarms for t...

Страница 22: ...the past data By pressing the DISPLAY key you can switch between one screen and two screens In the two screen setting the past data is displayed on the left side and the current data is on the right...

Страница 23: ...ENTER key When the Device Access Indicator is red the SD memory card is being accessed If you press the ENTER key while the card is being accessed the data may not be recorded properly Be sure to chec...

Страница 24: ...ulse input display is shown below 5 Where are the results of statistical operations displayed The statistical calculation result will be displayed on the digital statistical calculation screen Press t...

Страница 25: ...tput cable B 513 If you select Logic 4 lines appear at the top of the screen and the digital signal is displayed as an on off waveform 8 On the startup screen after turning on the power the screen dis...

Страница 26: ...terval at which the digital filter is enabled 10 The measured voltage and temperature fluctuations are not stable while recording data with external sampling If the AC line filter is set to Disabled O...

Страница 27: ...Web server function performs processing according to requests from PC browsers In the Remote Control and Screen Display screens the browser requests screen data from the device according to the numbe...

Страница 28: ...orded unless sampling pulses are input from outside Also if External Sampling is set to On the Sampling Interval item will be grayed out 14 What is the update timing of the waveform display on the mai...

Страница 29: ...measurement GL240 checks burnout disconnection periodically This burnout check applies a voltage which affects another instrument connected in parallel In such a case set the burnout setting to Off 16...

Страница 30: ...set up direct EXCEL function The output data of direct EXCEL transfer includes statistical values and instantaneous values To transfer an instantaneous value set the output data to Instantaneous Value...

Страница 31: ...MTP port number to 587 Adjust the time zone to your region UTC 09 00 for Japan Set the SMTP authentication method to SMTP AUTH Set the SMTP AUTH to PLAIN Set the same as POP3 to No Set the SMTP user n...

Страница 32: ...ontroller Device Manager USB controller items If the USB driver is not installed correctly please refer to the USB Driver Installation Manual and install it Check the Recording Settings screen 1 Is th...

Страница 33: ...lt printer please restart the software And also restart the software if you have changed the settings of Default printer 7 When the temperature is measured with a thermocouple the temperature is displ...

Страница 34: ...GL840 APS stops receiving data However if the recording on the main unit is set to On the data is still recorded on the main unit 5 How to fix the problem when the wireless LAN unit option B 568 is in...

Страница 35: ...uipped with both a wireless LAN and a wired LAN and both are connected disabling the wired LAN may allow you to search correctly Go to Control Panel Network and Sharing Center Change Adapter Settings...

Страница 36: ...on the amount of free space on your PC s hard disk When the SD memory card in the main unit is full recording in the main unit will stop but data transfer to the PC will continue In the case of Off d...

Страница 37: ...ng setting of GL100_GL240_GL840 APS to Off 16 Is it possible to print out the data being played back in the scale it is displayed in data that spans multiple pages It is possible Specify the area to b...

Страница 38: ...on a large monitor There is no limit to the display size as long as the display resolution is 1024 x 768 or higher 20 Can I display the cursor as a horizontal cursor It is possible It is possible to s...

Отзывы: