5.4.1 Restoring factory calibration settings
If the thermometer value is entered before the bath temperature is completely stable the
calibration could be poor and liquid temperature readings will be incorrect. If the heating
circulator is not in accordance with the thermometer following calibration then it may not
have been successful and the equipment should be reset, using the restore factory
settings function.
1. Press the
F
and
S
buttons simultaneously for 3
seconds.
2. Rotate the dial to highlight
restore factory
, press
S
to select.
Yes/No are displayed.
3. Rotate the dial to toggle to
yes
, press
S
to select.
Factory calibration values are restored
.
4. Press
F
to return to the home screen
.
°c
Point 19.86 c
Calibrate Probe
Point 70.04 c
Add Point
Restore Factory
Yes